Analysis of Selected Anions with HPLC and Electrochemical Detection

By | May 14 2008

By Angelika Gratzfeld-Huesgen and Rainer Schuster

Traditionally, conductivity is the standard detection method for ion analysis. During the last 10 years, however, there has been increasing interest in electrochemical detection (ECD) for the analysis of anions. This is due to the high sensitivity and selectivity of this type of detection.

Figure 1 shows how certain anions can be selectively detected in the presence of several other anions. In total, 16 anions were injected. Four that can be oxidized: iodide, nitrite, bromide, thiocyanide, and twelve that cannot be oxidized: fluoride, chloride, azide, nitrate, phosphite, sulfate, molybdate, hydrogencarbonate, perchlorite, phosphate, selenate and perchlorate.


In order to demonstrate the selectivity and sensitivity, iodide in table salt was analyzed. There is currently a great deal of interest in the determination of iodide in various foodstuffs, because some data indicates that the consumption of iodide has increased, leading to a potential rise in thyroid disorders. Table salt is almost 100% sodium chloride and with conductivity detection a large chloride peak may overlap with the iodide. Figure 2 shows the selective detection of iodide in a large excess of cloride.

 

Detection potential is of great importance for selectivity and sensitivity and must be determined for each set of analysis parameters.

The Agilent 1049A ECD in combination with an auto-sampler provides a time-saving way to find the optimum potential for several compounds through its autoincrement mode. The potential is automatically increased within a series of runs and the best detection potential can be selected from the chromatogram plots.

An example is given in figure 3. The optimum potential was 1.1 V. At lower potentials the response was insufficient, at higher potentials the detector drift was unacceptable.

The minimum detectable level for iodide was about 40 μg iodide per kilogram table salt. The relative standard deviation for peak heights over 75 runs was 3% for 300 pg of iodide and an injection volume of 10 μl.

Source: Agilent

Sources

Agilent

Advertisement
Accelrys
Accelrys
Get Help with Your Lab Equipment
…or help others.

Build your Lab Bench and get notifications and updates for the lab equipment you own!

www.labwrench.com
  • Facebook
  • LinkedIn
  • Twitter
  • Vimeo
  • RSS
Now Available in the App Store!

Related Articles

Advertisement
W.S. Tyler
W.S. Tyler

Subscribe to the Magazine

Click Here for your free subscription to Lab Manager Magazine®

E Newsletters

Click Here to subscribe to any of the following Lab Manager Magazine® E Newsletters

  • Lab Manager Magazine Weekly Monitor
  • New Product Alerts
  • eAnalytical Monthly
  • eLifeScience Monthly
  • eLab Equipment Monthly
  • Product Surveys and Event Promotions
  • Events and Webinars
Sign Up for Newsletters

Lab Video

Advertisement
See All Lab Videos

Webinars & Events

See All Webinars & Events
  • Facebook
  • LinkedIn
  • Twitter
  • Vimeo
  • RSS
Now Available in the App Store!

Page Generated: May 23 2013

Advertisement