Lab Manager | Run Your Lab Like a Business

Product in Action

Increase Accuracy in Molar Mass Averages by Gel Permeation Chromatography

Gel permeation chromatography (GPC) is the most widely accepted and used analytical method for  obtaining molar mass averages of both synthetic and biopolymers. Traditionally, molar mass averages are obtained via a peak position calibration involving a series of standards of known molar mass and chemistry analyzed by GPC coupled to a differential refractive index (RI) detector.

by Tosoh Bioscience

Figure 1. EcoSEC GPC SystemIn the context of GPC, GPC/RI continues to be heavily employed as it provides excellent day-to-day reproducibility and is ideal for quality control procedures. However, a conventional RI detector, i.e. one with a stagnant reference side, typically has a large amount of RI baseline drift which has been shown to drastically affect the accuracy and precision of molar mass averages and distributions with errors up 25%. Here, we have studied the repeatability, reproducibility, and baseline stability of a dual-flow RI detector in the EcoSEC® GPC System for the determination of molar mass averages via peak position calibration. The dual flow design is shown to compensate for the changes in the refractive index of the solvent over time by continuously flowing pure solvent through the reference side of the flow cell.

Experimental

GPC analysis was performed on an EcoSEC GPC System equipped with a dual-flow refractive index detector or a modular HPLC system with an external conventional refractive index detector. Separation of polystyrene standards (PS) occurred over a column blank consisting of TSKgel® SuperMultiporeHZ-M columns, with THF as the mobile phase.

Results and Discussion

To demonstrate the repeatability, reproducibility, and baseline stability of a dual-flow RI detector compared to a conventional RI detector a series of identical experiments was performed on the EcoSEC GPC System (Figure 1) and two conventional HPLC systems. As shown in Figure 2, five consecutive injections of PS with run times of one hour without auto zeroing the detector between injections for a total of five hours, resulted in an extremely stable baseline with low baseline drift for the dual-flow RI detector and a significantly drifting baseline on the two conventional RI detectors.

Figure 2. Comparison of baseline drift of the dual-flow refractive index detector in the EcoSEC GPC System to that of two conventional refractive index detectorsThe repeatability and reproducibility of the molar mass averages as obtained via the dual-flow and conventional RI detectors were also compared. The reproducibility of the weight-average molar mass, Mw, of the dual-flow RI detector was determined to be superior by a factor of 3 to that of a conventional RI detector. Additionally, the day-to-day reproducibility and repeatability for the determination of molar mass averages was shown to vary less than 0.5% for the dual-flow RI detector, while the conventional RI detector produced day-to-day variations in molar mass averages between 1% and 3%.

Conclusion

A stable RI detector baseline is required for successful experiments, and repeatable and reproducible molar mass averages. Extreme care must be taken when molar mass averages and distributions are determined via peak position calibration as uncertainties and instabilities in the RI baseline can result in relatively large errors, inconsistencies, and deviations in molar mass averages. The repeatability and reproducibility of the molar mass averages were shown to increase greatly when a conventional RI detector, was replaced with a dual-flow RI detector.


Tosoh Bioscience LLC
3604 Horizon Drive, Suite 100, King of Prussia, PA 19406
tel. (484) 805-1219, fax (610) 272-3028
website: www.tosohbioscience.com