Anion Analysis by Ion Chromatography: EPA Method 300.1 in Practice

EPA Method 300.1 measures two sets of anions on the same column with the same eluent. The only difference between them is the injection: 10 µL for common anions, 50 µL for disinfection byproducts.

Written byTrevor J Henderson
| 5 min read
Water utility lab supervisor adding preservative to an amber sample bottle, with an ion chromatography system behind him.
Register for free to listen to this article
Listen with Speechify
0:00
5:00

Key takeaways

  • EPA Method 300.1 uses ion chromatography with suppressed conductivity to measure inorganic anions in reagent, surface, ground, and finished drinking water.
  • Part A covers common anions: fluoride, chloride, nitrite, bromide, nitrate, ortho-phosphate, and sulfate; Part B covers disinfection byproducts bromate, chlorite, and chlorate, plus bromide.
  • Both parts use the same columns and eluent; the only difference is injection volume, 10 µL for Part A and 50 µL for Part B on a 2 mm column.
  • Part B samples are preserved with 50 mg/L ethylenediamine (EDA), which protects chlorite, chlorate, bromate, and bromide; chlorite samples go in opaque bottles and are held cold for up to 14 days.
  • Batches of up to 20 field samples need calibration checks, blanks, fortified blanks and matrices, duplicates, and a dichloroacetate surrogate within 90 to 115% recovery.

 

Read the method text and one sentence stands out. EPA Method 300.1 divides its anions into Part A, the common anions, and Part B, the inorganic disinfection byproducts, and then states that the only difference between the two parts is the volume of sample analyzed. Columns, eluent, and suppressor are identical; Part A injects 10 µL and Part B injects 50 µL on a 2 mm column, because disinfection byproducts occur at far lower concentrations than chloride or sulfate.

What does EPA 300.1 measure? Inorganic anions in drinking water and its sources, using ion chromatography with suppressed conductivity: seven common anions in Part A and the disinfection byproducts bromate, chlorite, and chlorate, plus bromide, in Part B. It remains a routine method for drinking water compliance, and for high-volume labs, success comes down to preservation, QC discipline, and keeping the column healthy. This article covers each. For IC basics, see our ion chromatography overview.

What does EPA Method 300.1 measure?

PartAnalytesInjection (2 mm column)
Reporting Units

Part A: common anions

Fluoride, chloride, nitrite-N, bromide, nitrate-N, ortho-phosphate-P, sulfate

10 µL

mg/L

Part B: inorganic disinfection byproducts

Bromate, chlorite, chlorate, bromide

50 µL

µg/L

 

The method notes that bromide appears in both parts: in Part A as a common anion, and in Part B because it’s a critical precursor of disinfection byproducts. Nitrite and nitrate are reported as N, and phosphate as P. With a 4 mm column, injection volumes rise to 40 µL and 200 µL to compensate for the larger column cross-section.

How does 300.1 differ from 300.0?

EPA Method 300.0 (1993) also uses IC for anions, but with a broader scope. As the NEMI summary of 300.0 describes, its Part A covers drinking water, surface water, wastewater, groundwater, solids after extraction, and leachates, while Part B covers bromate, chlorate, and chlorite. Method 300.1 (1997) focuses on drinking water and its sources, and was developed around a high-capacity anion exchange column that tolerates high-ionic-strength samples, with a surrogate and EDA preservation built into the method. Check your permit or regulatory program to confirm which method applies.

Why do disinfection byproducts matter?

Bromate forms when ozone reacts with bromide; chlorite and chlorate come from chlorine dioxide and hypochlorite use. Under EPA’s disinfection byproducts rules, summarized in its plain-English guide, bromate and chlorite are regulated in drinking water. That makes Part B sensitivity central: the method’s single-laboratory detection limits for bromate and chlorite in reagent water are about 1.44 µg/L and 0.89 µg/L, as the NEMI summary of 300.1 lists. Organic DBP precursors are tracked with TOC; see our TOC analyzer guide.

Interested in chemistry and materials science?

Register for a FREE Lab Manager account to subscribe to our Chemistry & Materials Science Newsletter.
Subscribe for Free

Sample collection and preservation

AnalytePreservationHolding Time

Bromide, chloride, fluoride

None required

28 days

Nitrate-N, nitrite-N, ortho-phosphate-P

Cool to 4 °C

48 hours

Sulfate

Cool to 4 °C

28 days

Bromate, chlorate

50 mg/L EDA

28 days

Chlorite

50 mg/L EDA, cool to 4 °C, opaque bottle

14 days

 

  • Use EDA for Part B. Add EDA to a final 50 mg/L, equivalent to 0.5 mL of a 100 mg/mL solution per liter. EDA chelates iron that catalyzes chlorite breakdown and ties up hypochlorite, protecting chlorite, chlorate, and bromate.
  • Protect bromide. In the method’s stability study, bromide in unpreserved chlorinated drinking water fell to about 9% recovery by day 0, while EDA-preserved samples held near 100%.
  • Sparge chlorine dioxide samples. Residual chlorine dioxide forms more chlorite; purge with inert gas for about five minutes before adding EDA, at collection.
  • Keep chlorite in the dark in opaque or amber bottles.
  • Watch phosphate. Samples for ortho-phosphate shouldn’t sit at room temperature more than 12 cumulative hours.
  • Add EDA to standards and QC too. The method requires EDA in Part B calibration standards, blanks, and fortified blanks to normalize any EDA-related bias.

QC requirements

QC element
Requirement

Analysis batch

Up to 20 field samples, with initial and end calibration checks, a laboratory reagent blank, and a laboratory fortified blank

Continuing calibration check

After every tenth field sample; Part B initial check must use the lowest calibration standard

Calibration check limits

75–125% from MRL to 10×MRL; 85–115% above

Fortified matrix (LFM)

At least 10% of samples; 75–125% recovery until in-house limits are established

Duplicates

At least 10% of samples or one per batch; RPD within ±20% (MRL to 10×MRL) or ±10% (higher)

Surrogate (dichloroacetate)

Added to every sample and standard; 90–115% recovery

Peak shape

Peak Gaussian Factor of the surrogate between 0.80 and 1.15

Retention time

Calibration checks fail if retention shifts more than ±5%; surrogate shifts above 2% signal a system problem

Quality control sample

External source, at start and quarterly, within ±15%

MDLs

Determined over at least three days and repeated every six months

 

All requirements are from the method text. State laboratory SOPs, such as Georgia EPD’s 300.1 procedure for bromate and chlorite, add program-specific details on top.

Throughput tips for high-volume labs

  • Plan around a 25-minute run. The method recommends 25 minutes to clear late-eluting peaks seen in ozonated and chlorine dioxide matrices; don’t shorten it without checking for carryover.
  • Run Part A and Part B on dedicated systems or loops to avoid constant changeovers, or schedule them in blocks.
  • Trend retention times. The method requires a historic record and suggests cleaning or replacing the column when retention falls below 80% of its original value.
  • Validate modern options. The method allows different columns, injection volumes, or eluents, including generated hydroxide eluents, if the initial demonstration and baseline criteria are repeated.
  • Be careful diluting eluent. Diluting carbonate eluent can put a bicarbonate peak in the bromate window.
  • Keep water and reagents clean. Contaminated reagent water raises baselines and causes false positives.
  • Filter samples at 0.45 µm with filters shown to be free of ionic contaminants.

For confusion between IC and protein ion exchange, see our explainer on ion chromatography vs. ion exchange chromatography. For EPA metals digestion methods, see our guide to EPA 3051A and 3052. System maintenance is covered in IC System Maintenance: Suppressors, Eluent Generators, and Column Life.

An EPA 300.1 readiness checklist

  1. The right method is confirmed. Check whether your program specifies 300.0 or 300.1, and which analytes and reporting levels apply.
  2. Sample kits match the analytes. Opaque bottles, EDA, coolers, and sparging gear for chlorine dioxide plants should go out with samplers.
  3. Holding times are tracked from collection. Nitrite, nitrate, and phosphate have 48-hour limits; chlorite has 14 days with preservation.
  4. EDA is added consistently. Field samples, standards, blanks, and fortified samples for Part B all need EDA at 50 mg/L.
  5. Batches meet QC structure. Calibration checks, blanks, fortified samples, duplicates, and surrogate recovery are required in every batch of up to 20 samples.
  6. Column health is trended. Retention time and surrogate peak shape records show when to clean or replace columns.
  7. Method changes are re-validated. New columns, eluents, or injection volumes require repeating the initial demonstration of performance.
  8. Detection limits are current. Repeat MDLs every six months and when operators or instrument response change.

EPA 300.1 is a mature method, but its details still decide data quality: an EDA addition at the tap, a 48-hour clock on nitrate, a surrogate peak that’s drifting. Build those details into sample kits, batch templates, and trend charts, and the method will run reliably at high volume.

 

This article was produced under Lab Manager’s AI Editorial Guidelines

Frequently Asked Questions (FAQs)

  • What does EPA Method 300.1 measure?

    Inorganic anions in reagent, surface, ground, and finished drinking water: fluoride, chloride, nitrite, bromide, nitrate, ortho-phosphate, and sulfate in Part A, and the disinfection byproducts bromate, chlorite, and chlorate, plus bromide, in Part B.

  • What is the difference between EPA 300.0 and 300.1?

    Both use ion chromatography for anions. EPA 300.0 has a broader scope including wastewater and solids, while 300.1 focuses on drinking water and its sources, with a high-capacity column, a surrogate, and EDA preservation for disinfection byproducts.

  • How are samples preserved for EPA 300.1?

    Part B samples are preserved with 50 mg/L ethylenediamine. Chlorite samples also go in opaque bottles and are cooled to 4 °C. Nitrate, nitrite, phosphate, and sulfate samples are cooled to 4 °C.

  • What is the holding time for chlorite in EPA 300.1?

    14 days when preserved with 50 mg/L EDA and cooled to 4 °C.

  • What surrogate is used in EPA 300.1?

    Dichloroacetate, added to every sample and standard, with required recovery of 90% to 115%.

Add Lab Manager as a preferred source on Google

Add Lab Manager as a preferred Google source to see more of our trusted coverage.

About the Author

  • Trevor Henderson headshot

    Trevor Henderson BSc (HK), MSc, PhD (c), has more than two decades of experience in the fields of scientific and technical writing, editing, and creative content creation. With academic training in the areas of human biology, physical anthropology, and community health, he has a broad skill set of both laboratory and analytical skills. Since 2013, he has been working with LabX Media Group developing content solutions that engage and inform scientists and laboratorians. He can be reached at thenderson@labmanager.com.

    View Full Profile

Related Topics

Related Articles

Current Magazine Issue Background Image

CURRENT ISSUE - October 2026

Why Lab Teams Resist Change—And How Leaders Can Respond

Addressing the human side of automation, AI, and operational change

October 2026 Lab Manager Cover Image