Beyond the AST Result: What Lab Leaders Need to Get Right

Reliable antimicrobial susceptibility testing depends on more than technical accuracy. Clear training, competency assessments, and review procedures help labs maintain consistent results and support clinical decisions

Written byFarah Tayyab, MLS(ASCPi), M(ASCPi), CLT(NYSED)
| 4 min read
Antimicrobial susceptibility test
Register for free to listen to this article
Listen with Speechify
0:00
4:00

Antimicrobial susceptibility testing (AST) is often viewed as a technical process: identify the organism, perform susceptibility testing, verify the result, and report it. But each AST result can influence more than an individual laboratory report. It can help guide patient treatment and later contribute to cumulative susceptibility data used for antimicrobial stewardship and other clinical decisions. Even minor gaps in training, quality control, interpretation, or result review can therefore have consequences beyond a single test.

After more than 10 years in clinical microbiology, I have learned that a strong AST program is built around the entire testing process. Technology is important, but it works best when staff understand the method, quality systems support the work, and there is a clear plan for reviewing results that do not fit the expected pattern.

Training should cover more than patient testing

When a laboratory introduces a new AST method, training should cover the entire workflow, including quality control, reagent and material storage, instrument operation, maintenance, result review, and troubleshooting.

This is especially important with manual methods. In gradient diffusion MIC testing, factors such as strip storage and handling, inoculum preparation, incubation conditions, and endpoint reading can affect performance. Knowing the basic steps is not enough if important details that protect result quality are overlooked.

Implementation should be treated as a process rather than a single training session. CLSI M52 provides recommendations for verification and post-verification quality assurance of commercial identification and AST systems. A new method should move into routine use only after the laboratory has completed the required verification and confirmed that staff are trained to perform and review the testing correctly. 

Refresher training can reinforce steps that continue to affect test quality, such as thoroughly vortexing inoculum suspensions and confirming that turbidity meets the required McFarland standard. The goal is to maintain consistency while reminding staff why these details matter.

Changes after implementation also require clear communication. Software or instrument updates, new manufacturer instructions, and procedure revisions may affect testing or result review. Focused training can help staff understand what changed and what they need to do differently.

Breakpoint updates require the same coordinated approach. Laboratories should designate who is responsible for monitoring new breakpoint recommendations and who has authority to approve implementation. Before a change goes live, the lab should confirm that the testing system and laboratory information system can support the updated breakpoints, complete any required verification, and update procedures and reporting rules. Leadership should also identify who is responsible for each step and communicate the change to affected staff so that instrument settings, LIS configuration, procedures, and reporting remain aligned.

Competency should reflect everyday practice

The way staff competency is assessed matters as much as completing the assessment itself. One issue I have seen is that staff naturally pay extra attention when they know they are working on a competency sample. That may show they can perform the procedure correctly, but it may not fully reflect everyday practice.

One practical approach I have encountered is blind competency testing, in which the technologist does not know in advance that a sample is being used for assessment. In some situations, the same material can also be tested independently by two staff members. Comparing results can help identify differences in technique, interpretation, or workflow that may otherwise be missed.

Lab manager academy logo

Lab Quality Management Certificate

The Lab Quality Management certificate is more than training—it’s a professional advantage.

Gain critical skills and IACET-approved CEUs that make a measurable difference.

These approaches do not replace an organization's required competency program. They can, however, give leaders additional information about routine performance. The goal is not to catch someone making a mistake. It is to find gaps early, provide training, clarify procedures, and improve consistency across the team.

Give staff a pathway for unusual results

Unusual susceptibility patterns are where technical knowledge and laboratory systems come together. An unexpected result should not be accepted simply because an instrument generated it, but it also should not be rejected only because it looks unusual. Staff need a consistent way to investigate what they are seeing.

In my experience, the review starts with the basics. Are recent QC results acceptable? Is instrument maintenance current? Were required cleaning procedures completed? Does the organism identification and phenotype make sense with the susceptibility pattern? Could there be a technical explanation? Depending on the organism and finding, additional phenotypic testing or an appropriate molecular method may help clarify the result before it is finalized. CLSI M100 also provides guidance for interpreting and reporting susceptibility results, including certain unusual results that may require further review before reporting.

Interested in lab leadership?

Register for a FREE Lab Manager account to subscribe to our Lab Leadership Digest Newsletter.
Subscribe for Free

When an unusual result appears, the next steps should be clear and consistent, regardless of which staff member is working. Laboratories benefit from clear procedures that tell staff what to review, when repeat or additional testing is appropriate, and when a supervisor, pathologist, or other expert should be involved.

Remember where the data go next

Each AST result can become part of a larger picture. These results are combined in antibiograms to show local susceptibility patterns and support antimicrobial stewardship. This is why accurate AST results matter not only for individual patients, but also for understanding resistance patterns in the laboratory.

Working with antibiogram data taught me to pay close attention to the denominator. A susceptibility percentage can look convincing until you see how few isolates produced it. With a small number of isolates, only a few results can noticeably change the percentage. CLSI M39 provides standardized guidance for collecting, analyzing, and presenting cumulative susceptibility data for clinical decision-making and stewardship.

Leaders should therefore look beyond the final percentage. They need to understand how isolates were selected, whether data were extracted consistently, whether testing or breakpoints changed during the period being compared, and whether the isolate count supports the interpretation. A polished report is only as dependable as the data behind it.

Reliable AST is a leadership responsibility

AST quality is sometimes discussed as though it belongs mainly at the bench. In reality, laboratory leaders shape many of the conditions that determine whether results can be trusted. They influence how a new method is introduced, how staff are trained, how competency is assessed, how breakpoint changes are implemented, how quality problems are investigated, and how data are reviewed after testing is complete.

The most effective approach is to make the important steps clear: train staff on the whole method, communicate and coordinate changes before they become routine, assess performance in a way that reflects everyday work; and give the team a defined pathway for unusual results. When AST data are used for larger decisions, make sure the numbers are understood in context.

For laboratory leaders, getting these details right builds confidence in the results that clinicians, stewardship teams, infection prevention professionals, and other laboratory partners rely on.

Frequently Asked Questions (FAQs)

  • What is antimicrobial susceptibility testing (AST)?

    AST is a process used to determine the susceptibility of microorganisms to antimicrobial agents, guiding patient treatment and informing clinical decisions regarding antimicrobial stewardship.

  • What role do laboratory leaders play in ensuring reliable AST results?

    Laboratory leaders are responsible for shaping conditions that impact AST quality, including staff training, assessing performance, implementing changes, and ensuring clear communication about results.

Add Lab Manager as a preferred source on Google

Add Lab Manager as a preferred Google source to see more of our trusted coverage.

About the Author

  • Placeholder image

    Farah Tayyab is a specialty laboratory technologist in the microbiology laboratory at Mount Sinai Hospital in New York. She has more than 10 years of experience in clinical microbiology, antimicrobial susceptibility testing, molecular diagnostics, resistance-pattern analysis, laboratory quality systems, and cumulative susceptibility data.

    View Full Profile

Related Topics

Loading Next Article...
Loading Next Article...
Current Magazine Issue Background Image

CURRENT ISSUE - October 2026

Why Lab Teams Resist Change—And How Leaders Can Respond

Addressing the human side of automation, AI, and operational change

October 2026 Lab Manager Cover Image