Class II vs. Class III: Matching Biological Safety Cabinets to Biosafety Levels

A practical comparison of Class II and Class III containment equipment, with guidance on selecting the right cabinet for each biosafety level

Written byErika Russell
| 4 min read
A female laboratory technician wearing a lab coat and mask uses a multichannel pipette inside a Class II Type A2 biological safety cabinet.
Register for free to listen to this article
Listen with Speechify
0:00
4:00

Selecting the correct containment equipment is one of the most consequential decisions a laboratory can make, and biological safety cabinets sit at the center of that decision. The choice between Class II and Class III cabinets depends on the biosafety level (BSL) of the work, the risk group of the agent in use, and the procedures performed. According to the Biosafety in Microbiological and Biomedical Laboratories (BMBL), 6th edition, published by the CDC and NIH, cabinet selection should reflect both the agent's hazard classification and the protection needed for personnel, product, and environment.

What biological safety cabinets are and how they differ from fume hoods

Biological safety cabinets are ventilated enclosures designed to protect personnel, the environment, and research materials from exposure to infectious aerosols and splashes. They use high-efficiency particulate air (HEPA) filtration to capture particles 0.3 micrometers in diameter with at least 99.97 percent efficiency, as defined by NSF/ANSI 49, the standard governing Class II cabinet design and certification.

Chemical fume hoods, by contrast, are not designed for biological containment because they exhaust unfiltered air and rely on directional airflow alone. A fume hood should never substitute for a biological safety cabinet when working with infectious agents.

The three primary classes of biological safety cabinets are Class I, Class II, and Class III. Class I cabinets protect personnel and the environment but not the product, while Class II and Class III cabinets provide all three forms of protection at differing levels of containment.

How biosafety levels guide cabinet selection

Biosafety levels are tiered designations (BSL-1 through BSL-4) that describe the containment practices, safety equipment, and facility design required for handling biological agents of increasing risk. The BMBL aligns each level with specific cabinet recommendations based on transmission routes, agent virulence, and procedural risk.

The general alignment is as follows:

Biosafety levelTypical agentsRecommended cabinet
BSL-1Non-pathogenic agents (e.g., non-pathogenic E. coli strains)Cabinet not required
BSL-2Agents with moderate hazard (e.g., Staphylococcus aureus, HIV in clinical specimens)Class II (A2, B1, B2)
BSL-3Aerosol-transmissible agents (e.g., Mycobacterium tuberculosis)Class II (B1, B2) or Class III
BSL-4High-consequence agents (e.g., Ebola virus, Marburg virus)Class III or Class II in positive-pressure suit lab

The selection also depends on whether volatile chemicals or radionuclides are used, since only certain Class II subtypes (B1 and B2) and Class III cabinets are hard-ducted to exhaust such materials safely.

What defines a Class II biological safety cabinet

A Class II biological safety cabinet provides personnel, product, and environmental protection through a combination of inward airflow at the work opening and HEPA-filtered downward airflow across the work surface. The inflow velocity prevents aerosols from escaping, while the laminar downflow protects the sample from contamination.

Lab manager academy logo

Lab Quality Management Certificate

The Lab Quality Management certificate is more than training—it’s a professional advantage.

Gain critical skills and IACET-approved CEUs that make a measurable difference.

Class II cabinets are divided into four subtypes (A1, A2, B1, and B2), distinguished by airflow patterns, exhaust configuration, and suitability for volatile chemicals. The differences are summarized below.

SubtypeMinimum inflow velocityExhaustVolatile chemical use
A175 ft/minRecirculated or exhausted to roomNot permitted
A2100 ft/minRecirculated or canopy-exhaustedMinute quantities only
B1100 ft/minHard-ducted, partial exhaustMinute quantities in direct exhaust zone
B2100 ft/minHard-ducted, 100 percent exhaustPermitted

Type A2 is the most common configuration in clinical and research laboratories because it balances containment performance with installation flexibility. Class II cabinets are appropriate for BSL-2 and BSL-3 work when paired with appropriate facility controls and personal protective equipment, according to the World Health Organization Laboratory Biosafety Manual, 4th edition.

What defines a Class III biological safety cabinet

A Class III biological safety cabinet is a gas-tight, fully enclosed, negative-pressure containment unit operated through attached rubber gloves, providing the highest level of personnel protection available in a cabinet. Often called a glove box, it is designed for work with the most hazardous pathogens, including Risk Group 4 agents.

Air entering a Class III cabinet passes through one HEPA filter, and exhaust air passes through two HEPA filters in series or one HEPA filter and an incinerator before discharge. The cabinet maintains a minimum negative pressure of 0.5 inches of water gauge relative to the surrounding room, ensuring that any breach results in inward, not outward, airflow.

Interested in lab tools and techniques?

Register for a FREE Lab Manager account to subscribe to our Lab Tools & Techniques Newsletter.
Subscribe for Free

Class III cabinets are typically interconnected with autoclaves and pass-through chambers so that materials can enter and exit without breaking containment. They are required for BSL-4 laboratories that do not use positive-pressure protective suits and are the only cabinet class suitable for manipulation of agents such as variola virus or Ebola virus in non-suit environments.

How to verify biological safety cabinet performance and certification

Biological safety cabinets must be certified upon installation, after any relocation or repair, and at least annually thereafter. Certification confirms that the cabinet meets the performance criteria established by NSF/ANSI 49 for Class II cabinets or by manufacturer and institutional specifications for Class III cabinets.

Required tests typically include:

  • HEPA filter integrity (DOP or PAO leak testing)
  • Inflow and downflow velocity measurements
  • Airflow smoke pattern visualization
  • Site installation assessment for vibration, lighting, and noise
  • Electrical safety verification

Only technicians accredited by the NSF International Field Certifier Accreditation Program should perform Class II certification. Records of certification should be retained according to institutional biosafety policies and made available for regulatory inspection.

Why procedural risk shapes biological safety cabinet selection

Matching a biological safety cabinet to a biosafety level requires evaluating procedural risk in addition to the agent's classification. Two laboratories handling the same organism may need different cabinets if one performs high-aerosol procedures such as sonication, vortexing, or centrifugation of open tubes, while the other limits work to closed-system manipulations. The BMBL emphasizes that aerosol-generating procedures with BSL-2 agents may warrant BSL-3 practices and equipment, including a Class II Type B cabinet with hard-ducted exhaust. Similarly, work with BSL-3 agents in large volumes or high concentrations may justify a Class III cabinet even when Class II would suffice for smaller-scale manipulations. Risk assessment, documented in writing and reviewed by the institutional biosafety committee, is the foundation of appropriate cabinet selection and should be revisited whenever protocols, agents, or personnel change.

Choosing biological safety cabinets through evidence-based risk assessment

Choosing between Class II and Class III biological safety cabinets is ultimately a function of the agent, the procedure, and the facility's containment infrastructure. Class II cabinets serve the vast majority of clinical and research laboratories operating at BSL-2 and BSL-3, while Class III cabinets are reserved for the most hazardous work, typically at BSL-4. Reliable selection depends on documented risk assessment, adherence to the BMBL and WHO guidance, and rigorous certification of biological safety cabinets to ensure they perform as designed.

This content includes text that has been generated with the assistance of AI. For more information, view Lab Manager’s AI use policy.

Add Lab Manager as a preferred source on Google

Add Lab Manager as a preferred Google source to see more of our trusted coverage.

Frequently Asked Questions (FAQs)

  • What is the main difference between Class II and Class III biological safety cabinets?

    Class II cabinets have an open front with inward airflow, while Class III cabinets are gas-tight, fully enclosed glove boxes operated under negative pressure for the highest level of containment.

  • When should a Class II Type B2 cabinet be selected over a Type A2?

    A Type B2 cabinet should be selected when work involves volatile toxic chemicals or radionuclides, because it exhausts 100 percent of air through hard ducting to the outside rather than recirculating it.

  • Why is annual certification of biological safety cabinets required?

    Annual certification verifies HEPA filter integrity, airflow velocities, and containment performance, ensuring the cabinet continues to protect personnel, product, and environment as specified by NSF/ANSI 49.

  • How does biosafety level determine which cabinet to use?

    The biosafety level reflects the agent's risk and the procedures performed, with BSL-2 and BSL-3 typically using Class II cabinets and BSL-4 requiring Class III cabinets or Class II cabinets within a positive-pressure suit laboratory.