Controlling Blanks in Trace Metals Sample Prep for ICP-MS

At parts-per-trillion levels, a single speck of dust can outweigh the analyte you’re measuring. In ICP-MS sample prep, the lab itself is usually the biggest source of contamination, and blanks are how you find it.

Written byTrevor J Henderson
| 5 min read
Trace metals analyst pipetting into fluoropolymer vials inside a laminar flow clean bench beside a sub-boiling still.
Register for free to listen to this article
Listen with Speechify
0:00
5:00

Key takeaways

  • In ICP-MS sample preparation, contamination usually comes from the lab: reagents, labware, digestion vessels, airborne particles, and people.
  • Procedural (method) blanks, carried through every step of digestion and dilution, are the main tool for detecting contamination; if they exceed the detection limit, the source must be found before data are reported.
  • Use high-purity or sub-boiled acids, 18 MΩ·cm water, fluoropolymer or polypropylene labware instead of glass, and acid-leach new labware.
  • Control the environment with HEPA-filtered clean benches or clean rooms, powder-free gloves, dedicated pipettes, and separation of high- and low-level work.
  • Diagnose high blanks step by step, from instrument to reagents to labware to the full procedure, and track blank trends over time.

 

Suppose a 10 µg dust particle contains 0.1% zinc. That’s about 10 ng of zinc. Drop it into a 10 mL digest and it adds roughly 1,000 ng/L, or one part per billion, to the sample, which may be orders of magnitude more than the zinc you were trying to measure. At the levels ICP-MS can see, the lab itself becomes the most important variable.

Controlling ICP-MS sample preparation blanks means treating everything that touches the sample, acids, water, vessels, pipette tips, air, and the analyst, as a potential source of the elements you’re measuring, and using procedural blanks to prove each step is clean. As a widely cited 2010 review in Analytical and Bioanalytical Chemistry on contamination in multi-elemental trace analysis shows, sub-ng/L work is achievable, but only in a well-designed and managed laboratory. This article explains where blanks come from, how to reduce them, and how to find the source when they rise. For digestion systems, see our guide to microwave digestion systems.

Which blanks matter?

BlankWhat it ContainsWhy it Matters

Calibration (instrument) blank

Diluent acid run on the instrument

Instrument and rinse contamination or memory

Reagent blank

Acids and water used, without sample

Reagent purity

Procedural (method or preparation) blank

Reagents carried through every digestion and dilution step, without sample

Total contamination added by the workflow

Labware or leach blank

Dilute acid soaked in vials, tubes, or vessels

Contamination from specific labware

Field or trip blank

Clean matrix taken to the field and back

Contamination during sampling and transport

 

The procedural blank is the most important. As Lab Manager’s guide to sub-ppb ICP-MS accuracy notes, if a preparation blank contains analyte above the method detection limit, the contamination source must be identified and eliminated before sample data are reported. EPA Method 3051A sets a similar rule for reagents: the reagent blank must be less than the lower level of quantitation for a reagent to be used.

Where does contamination come from?

SourceExampleWhy it Matters

Water

Deionized water supply, storage bottles

The largest component of most samples

Acids

Nitric, hydrochloric, HF, peroxide

Concentrated reagents carry trace metals into every sample

Labware

Glassware, volumetric flasks, caps, sample tubes, pipette tips

Glass leaches sodium, boron, and silicon; plastics carry manufacturing residues

Digestion vessels

Fluoropolymer liners, caps

Memory from previous high-level samples

Laboratory air

Dust and particles

Particles carry common metals onto open samples

People

Skin, hair, dust from clothing, gloves, cosmetics

Shed material carries many trace elements

Instrument

Sample introduction, cones, autosampler tubing

Memory and carryover between samples

 

A Thermo Fisher guide to ICP-MS contamination identifies labware and reagents as the biggest sources during sample preparation, naming pipette tips, volumetric flasks and caps, sample tubes, deionized water supplies, and nitric acid, and adds that hair, skin cells, and dust from clothing are significant sources too. Lab Manager’s sub-ppb guide notes that borosilicate glass can leach sodium, boron, and silicon into acidic samples.

How do you reduce contamination in ICP-MS sample preparation?

Reagents and water

  • Use acid grades suited to your detection limits, and check each lot’s certificate of analysis for elemental impurities.
  • Consider sub-boiling distillation of acids in fluoropolymer or quartz stills; the 2010 review describes it as a simple way to produce high-purity acids.
  • Use 18 MΩ·cm water from a well-maintained purification system, dispensed fresh.
  • Pour acids into clean secondary containers rather than pipetting from stock bottles, to keep stock clean.

Reagent purity drives blanks in the same way solvent grade drives chromatography baselines; see our guide to choosing solvent grades.

Labware and vessels

  • Replace glass with fluoropolymer (PFA) or polypropylene wherever possible.
  • Acid-leach new labware before first use to remove manufacturing residues.
  • Test pipette tips from different suppliers, or pre-rinse tips by drawing and discarding dilute nitric acid, as the Thermo guide suggests.
  • Clean digestion vessels between high- and low-level work; EPA 3051A specifies hot 1:1 hydrochloric acid, then hot 1:1 nitric acid, for at least two hours each; see our guide to EPA 3051A and 3052.
  • Keep separate labware sets for trace and high-level work.

Environment and people

  • Prepare samples in HEPA-filtered clean benches or clean rooms; cleanroom classification and design are covered in our cleanroom overview.
  • Wear powder-free nitrile gloves, lab coats dedicated to clean work, and hair covers.
  • Dedicate pipettes to trace work, as the sub-ppb guide recommends.
  • Keep containers capped, and minimize the time samples sit open.
  • Separate high-concentration work, standards preparation, and low-level samples in time and space.

Some ultra-low-background programs validate every container before use; one study of surface contamination from dust leached labware in hydrochloric and nitric acids, then incubated dilute nitric acid in each container at 80 °C for at least 12 hours and analyzed it to confirm low backgrounds.

How do you diagnose a high blank?

  1. Run a calibration blank to rule out instrument memory or rinse problems.
  2. Run reagent blanks from each acid and water source, including freshly opened lots.
  3. Leach-test suspect labware: tubes, tips, flasks, and vessels.
  4. Run procedural blanks with one variable changed at a time, such as a different vessel set or clean bench.
  5. Look at the element pattern: sodium, boron, and silicon suggest glass; elements common in dust suggest airborne particles; elements from previous high-level samples suggest vessel or instrument memory.
  6. Fix the source, then confirm with repeat procedural blanks before reporting data.

Setting and tracking blank limits

  • Set acceptance criteria relative to detection and quantitation limits, as methods and regulations require.
  • Run procedural blanks in every batch, and more often for ultra-trace work.
  • Trend blank results over time and by element; gradual increases often point to aging water systems, labware, or vessels.
  • Document corrective actions when blanks fail.
  • For pharmaceutical elemental impurities testing, low blanks are essential to meet sensitivity targets; see our guide to ICH Q3D and USP <233> sample prep.

A trace metals blank control checklist

  1. Procedural blanks run with every batch. They capture contamination from the whole workflow. If they exceed the detection limit, stop and find the source before reporting.
  2. Reagents are matched to detection limits. Check certificates of analysis for each acid lot; consider sub-boiling distillation for ultra-trace work.
  3. Water quality is verified. Use fresh 18 MΩ·cm water and monitor the purification system, since water is the largest component of most samples.
  4. Glass is replaced with fluoropolymer or polypropylene. Glass leaches sodium, boron, and silicon. Where glass is unavoidable, validate it with leach blanks.
  5. New labware is acid-leached and tested. Manufacturing residues and supplier variability can surprise you; leach-test before use.
  6. Digestion vessels are cleaned and segregated. Hot acid cleaning between high- and low-level samples prevents memory effects.
  7. Samples are prepared in clean air. HEPA-filtered clean benches or rooms, capped containers, and minimal open time reduce particle contamination.
  8. Personnel practices support clean work. Powder-free gloves, dedicated coats and pipettes, and hair covers limit human contamination.
  9. Blank trends are reviewed. Trending reveals slow drifts from aging equipment or supplies before they become failures.

At the detection limits modern ICP-MS can reach, every surface, reagent, and breath of air is part of the measurement. Treat the procedural blank as your most important sample, control what touches the digest, and investigate methodically whenever the blank moves.

 

This article was produced under Lab Manager’s AI Editorial Guidelines

Frequently Asked Questions (FAQs)

  • How do you reduce contamination in ICP-MS sample preparation?

    Use high-purity or sub-boiled acids and 18 MΩ·cm water, replace glass with fluoropolymer or polypropylene labware, acid-leach new labware, clean digestion vessels thoroughly, work in HEPA-filtered clean benches, wear powder-free gloves, and dedicate pipettes and labware to trace work.

  • What is a procedural blank in ICP-MS?v

    A procedural, method, or preparation blank contains only the reagents but is carried through every digestion and dilution step, so it captures all contamination added by the workflow.

  • Why should glass be avoided in trace metals analysis?

    Borosilicate glass can leach elements such as sodium, boron, and silicon into acidic samples, raising blanks. Fluoropolymer and polypropylene labware are preferred.

  • What should you do if an ICP-MS blank is high?

    Isolate the source step by step: check instrument blanks, then reagent blanks, then labware leach tests, then procedural blanks with one variable changed at a time. Fix the source and confirm clean blanks before reporting data.

  • What acid grade is needed for ICP-MS?

    It depends on detection limits. Trace metal grade acids are common, and sub-boiled or ultra-high-purity acids are used for ultra-trace work. Check each lot’s certificate of analysis and confirm with reagent blanks.

Add Lab Manager as a preferred source on Google

Add Lab Manager as a preferred Google source to see more of our trusted coverage.

About the Author

  • Trevor Henderson headshot

    Trevor Henderson BSc (HK), MSc, PhD (c), has more than two decades of experience in the fields of scientific and technical writing, editing, and creative content creation. With academic training in the areas of human biology, physical anthropology, and community health, he has a broad skill set of both laboratory and analytical skills. Since 2013, he has been working with LabX Media Group developing content solutions that engage and inform scientists and laboratorians. He can be reached at thenderson@labmanager.com.

    View Full Profile