A supplier’s certificate of analysis (CoA) often looks conclusive. There is a laboratory name, a batch number, perhaps a chromatogram, and a large purity figure near the top. It is tempting to file it and move on.
But a CoA is a summary of evidence, not a verdict. How much weight it deserves depends on two things: whether the testing can be tied to the material that actually arrived, and whether the methods used can support the conclusions being drawn from them. The five checks below work through both. Several borrow principles from pharmaceutical guidelines that do not apply to most research materials or suppliers, but that capture careful thinking about what analytical evidence can and cannot show.
1. Confirm the document belongs to the batch you received
Establish traceability before reading any result. The material name, lot number, certificate identifier, and date of analysis or release should match the container and its paperwork. If a separate testing laboratory is cited, its link to the batch should be explicit.
Be wary of documents that are not batch-specific: “typical” or “representative” certificates, reports without a lot number, or one certificate reused across different batches or shipments without a clear batch-specific link. An analysis date that falls before the stated manufacture date also deserves a question.
ICH Q7, the good manufacturing practice guide for active pharmaceutical ingredients, offers a useful template. It expects each incoming batch to carry a distinctive code, and it describes CoAs that give the material name, batch number, and release date; list each test with its acceptance limits and numerical results; and are dated and signed by authorized quality personnel.1 Research suppliers are not bound by Q7, but a certificate missing most of these elements is hard to rely on.
An excellent result proves little if you cannot show that the tested sample came from your batch. If the identifiers do not line up, place the material on hold, document the discrepancy, and contact the supplier for clarification or corrected batch-specific documentation. Do not release it for use until the link between the certificate and the material has been resolved.
2. Establish who performed the testing
“Third-party tested” means something only when the third party can be identified. Look for the testing laboratory’s name and address, a report number, the test date, and the methods used. A supplier-branded summary of an external report is not the report itself, so ask for the original. Where authenticity is uncertain, verify the report directly with the testing laboratory.
If accreditation is claimed, check it. ISO/IEC 17025 specifies requirements for the competence, impartiality, and consistent operation of testing and calibration laboratories.2 Accreditation against it is granted by accreditation bodies for a defined scope of tests, not for everything a laboratory does. The accreditation body can confirm a laboratory’s current status and scope, including whether the test on your certificate falls within it; the International Laboratory Accreditation Cooperation (ILAC) lists the signatory bodies that hold these records.3
Accreditation is not mandatory for every test, and unaccredited laboratories can produce sound data. But an accreditation claim that fails this check says something about the document as a whole.
3. Ask what each method can actually show
Trouble starts when a result from one technique is read as proof of something it was never designed to establish.
High-performance liquid chromatography (HPLC) separates the components of a sample and is well suited to impurity profiling. ICH Q6A states that identification solely by a single chromatographic retention time is not regarded as specific, whereas two chromatographic procedures based on different separation principles, or combined techniques such as HPLC with diode-array detection or HPLC–mass spectrometry (MS), are generally acceptable.4 Q6A is written for chemical drug substances and excludes some product classes from its scope, but the reasoning carries over.
A measured mass that matches the expected value is good supporting evidence of identity, but it is not proof. Isomers, and some structurally similar compounds, share the same nominal mass; telling them apart may require fragmentation data, chromatographic separation, or another orthogonal technique. A mass spectrum alone also says little about how much of the compound is present.
ICH Q2(R2), the current validation guideline, generalizes the point: where one procedure does not discriminate sufficiently, a combination of two or more is recommended.5
For each result, note the question it actually answers. If identity rests on a single retention time, or on mass alone where isomeric impurities are plausible, it is less firmly established than the certificate suggests. For critical materials, consider a confirmatory identity test on receipt. ICH Q7 generally expects at least one identity test for each incoming batch, even when an evaluated supplier’s CoA is used in place of other testing, subject to specified exceptions.1
4. Look behind the headline purity figure
A large “99 percent purity” is easy to read and easy to overinterpret.
Where chromatographic purity is reported as an area percentage, the main peak’s area is expressed as a percentage of the total integrated peak area, subject to the method’s integration and reporting criteria.6 The figure describes only what the detector recorded under those conditions. Detector response varies between substances, so a weakly responding impurity is underrepresented; one not fully resolved from the main peak may be partly counted with it; and anything that is not detected, or does not elute during the run, is not counted at all.6 Water, inorganic impurities, and residual solvents are normally measured by separate tests,4 so a material can show high chromatographic purity yet contain appreciably less of the target compound by weight.
For a lab manager, that means a purity figure should not be treated as a pass/fail number in isolation. If the certificate gives only a headline percentage, omits the chromatogram or method details, or leaves relevant impurities outside the method’s scope, ask the supplier for the underlying data or additional documentation before relying on the result. If the material is critical to the work and the remaining uncertainty could affect the study, consider independent confirmatory testing.
Purity, assay, and content are therefore not interchangeable. Chromatographic purity describes the relative composition of detected components; assay describes how much of the substance is actually present, and ICH Q6A treats it as a separate test from impurities.4
Look for the chromatogram itself, not just the number; the detector and wavelength; enough method detail to judge whether relevant impurities could have been seen; plausible integration, with no unexplained excluded peaks; and a stated acceptance criterion. Q2(R2) rests on the principle that a procedure must be shown to be fit for its intended purpose.5 A bare figure asks you to take that on trust.
5. List what the certificate does not establish
The final check is often the most revealing: write down the conclusions the documentation does not support.
A chromatographic purity result does not, by itself, establish identity, quantity, sterility, endotoxin levels, stability, or the absence of contaminants the method was not designed to detect. Mass spectrometric data can support identity without settling purity or quantity. And no result on the certificate speaks to how the material was stored and shipped after sampling.
Then compare that list with the supplier’s claims on labels, websites, and correspondence. A CoA does not support statements it does not contain. If material is described as sterile, low in endotoxin, or stable at room temperature, each claim needs its own evidence.
Even under ICH Q7, reliance on supplier certificates is paired with supplier evaluation and regular checks of their reliability.1 For most laboratories, the practical equivalent is periodic independent testing, scaled to how critical the material is.
A five-question review
Before accepting supplier analytical documentation, ask:
- Does this document clearly correspond to the batch we received?
- Who performed the testing, and can that be verified?
- What does each method actually demonstrate?
- Is there enough method and result information to interpret the figures correctly?
- Which claims about the material remain unsupported?
Good documentation does not replace professional judgment; it gives that judgment something to work with. A certificate of analysis is most useful when it is treated not as a badge to accept, but as evidence to evaluate.
References
- International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use. ICH Q7: Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients. Step 4 version. November 10, 2000. Accessed September 24, 2026. https://database.ich.org/sites/default/files/Q7%20Guideline.pdf
- International Organization for Standardization; International Electrotechnical Commission. ISO/IEC 17025:2017. General Requirements for the Competence of Testing and Calibration Laboratories. ISO; 2017. Accessed September 24, 2026. https://www.iso.org/standard/66912.html
- International Laboratory Accreditation Cooperation. Accredited conformity assessment bodies. ILAC. Accessed September 24, 2026. https://ilac.org/ilac-mra-and-signatories/accredited-conformity-assessment-bodies/
- International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use. ICH Q6A: Specifications: Test Procedures and Acceptance Criteria for New Drug Substances and New Drug Products: Chemical Substances. Step 4 version. October 6, 1999. Accessed September 24, 2026. https://database.ich.org/sites/default/files/Q6A%20Guideline.pdf
- International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use. ICH Q2(R2): Validation of Analytical Procedures. Step 4 version. November 1, 2023. Accessed September 24, 2026. https://database.ich.org/sites/default/files/ICH_Q2%28R2%29_Guideline_2023_1130.pdf
- United States Pharmacopeia. General chapter <621> Chromatography. Stage 4 harmonized text; official December 1, 2022. US Pharmacopeial Convention. Accessed September 24, 2026. https://www.usp.org/sites/default/files/usp/document/harmonization/gen-chapter/harmonization-november-2021-m99380.pdf