A catalyst that looks uniform in bulk measurements can have surface regions that behave very differently. A new imaging method from the National University of Singapore maps these differences by recording individual reaction events while the catalyst operates.
The technique, called direct electrocatalytic conversion to light emission through nano-confinement, or DELINE, links a nonfluorescent catalytic reaction to a rapid follow-on reaction that generates a fluorescent signal. Recording where those signals appear produces a spatial map of relative activity across the catalyst surface.
Researchers demonstrated the method on gold and copper catalysts during electricity-driven reactions. Their findings were reported in Nature Catalysis and described in a National University of Singapore announcement.
Reveal activity hidden by averaged measurements
The team first examined water dissociation on thin gold plates. When water splits at the metal surface, it produces surface-bound hydrogen and hydroxide ions. A probe molecule reacts with the hydroxide to produce the fluorescent signal detected by DELINE. Control experiments indicated that the probe did not change the gold's catalytic activity.
Across 39 gold plates, the researchers compared 195 surface regions. Edges and side surfaces were generally more active than broad, flat faces, although activity also varied within the faces. The system can analyze about 100 particles in a single microscope view during a three-minute recording, according to the university.
Mapping individual events adds another layer to operando catalyst characterization, which examines a material under working conditions rather than relying only on measurements collected before and after a reaction.
Match catalyst design to operating conditions
Changing the applied voltage revealed a trade-off. Regions whose reactions accelerated more strongly as voltage increased tended to have lower baseline reaction rates. Simulations and additional imaging suggested that uneven sodium-ion accumulation near the gold altered surrounding water molecules. The local environment increased voltage sensitivity but could also crowd the surface and slow molecular movement.
Tests on copper reinforced the need to evaluate each system independently. During water dissociation and nitrate reduction, the broad copper faces were more active than the edges—the reverse of the general pattern observed on gold.
For laboratories adopting the method, probe chemistry requires careful validation for every new catalyst and reaction. Surface preparation, microscope alignment, electrochemical calibration, illumination, image-processing thresholds, and environmental conditions could all influence the apparent distribution of events. Bulk measurements remain necessary to determine whether local patterns translate into device-level performance.
The researchers are developing additional probes for other electrocatalytic reactions and hope to visualize processes inside operating batteries. DELINE therefore functions less as a universal catalyst test than as a way to ask a more precise question: which parts of a material perform best under the conditions in which it will actually operate?
Shared microscopy facilities would also need to decide how the specialized electrochemical setup fits routine scheduling. User training should cover not only imaging but also electrode preparation, fluid handling, probe stability, and the controls required to distinguish genuine reaction events from background fluorescence or image-processing artifacts.
This article was created with the assistance of Generative AI and has undergone editorial review before publishing.