Ion Chromatography: How It Works and When Your Lab Needs It

In 1975, three Dow Chemical scientists solved the problem that kept common ions out of routine chromatography: the eluent conducted more than the analytes. Their fix, the suppressor, still drives ion chromatography today.

Written byTrevor J Henderson
Updated | 5 min read
Lab technician loading sample vials into the autosampler of a dual-channel ion chromatography system.
Register for free to listen to this article
Listen with Speechify
0:00
5:00

Key takeaways

  • Ion chromatography (IC) separates ions on ion-exchange columns and usually measures them by suppressed conductivity, making it the standard technique for inorganic anions, cations, and small ionic species.
  • It’s used for drinking water and wastewater anions, cations such as ammonium and sodium, pharmaceutical counterions, food and beverage ions and carbohydrates, and ultrapure water in power and semiconductor plants.
  • The suppressor, introduced by Small, Stevens, and Bauman in 1975, lowers the eluent’s background conductivity and raises the analyte signal.
  • Unlike ICP, which measures total elements, IC identifies specific ionic species, such as nitrate versus nitrite or bromide versus bromate.
  • When choosing a system, consider anion and cation channels, eluent generation, detectors, sample preparation, automation, and maintenance.

 

Before 1975, measuring chloride, nitrate, and sulfate meant a separate wet chemistry test for almost every ion. Chromatography could separate ions, but detecting them was the problem: the eluent needed to push ions through an ion-exchange column was itself highly conductive, swamping the signal. That year, Hamish Small, T. S. Stevens, and W. C. Bauman of Dow Chemical published a method that added a second column, a suppressor, to strip away the eluent’s conductivity before the detector. As a National Bureau of Standards review put it, the concept paired ion exchange separation with an ion suppression system so analytes could be detected by conductivity without overwhelming background from the eluent.

Ion chromatography is the technique that grew from that idea: separation of ions on ion-exchange columns, usually followed by suppressed conductivity detection, to measure inorganic anions, cations, and small ionic organics in water, food, pharmaceuticals, and industrial materials. It’s the routine method for ions that other techniques handle poorly or can’t distinguish. This overview explains what IC is used for, how it works, and what to look for in a system.

What is ion chromatography used for?

ApplicationTypical Analytes

Drinking water and wastewater

Fluoride, chloride, nitrite, nitrate, bromide, sulfate, phosphate, bromate and other disinfection byproducts

Cation analysis

Lithium, sodium, ammonium, potassium, magnesium, calcium, amines

Pharmaceuticals

Counterions, inorganic impurities, ionic excipients

Food and beverage

Nitrate and nitrite, organic acids, sugars and other carbohydrates

Power plants and semiconductors

Trace anions and cations in ultrapure and process water

Environmental monitoring

Ions in rain, atmospheric particulates, soils, and ice cores

 

A Wiley reference chapter on IC in environmental analysis describes the technique as a subset of liquid chromatography for ionic solutes, including inorganic anions, cations, transition metals, and low-molecular-weight organic acids and bases, with ion exchange the primary separation mode and suppressed conductivity the primary detection method. A review of cation applications documents uses from drinking water and wastewater to rainwater and ice cores. Drinking water anion analysis is covered in Anion Analysis by Ion Chromatography: EPA Method 300.1 in Practice, and pharmaceutical uses in Ion Chromatography in Pharma QC: Counterions, Impurities, and Compendial Methods.

How does ion chromatography work?

  1. Eluent. A dilute ionic solution, such as carbonate/bicarbonate or potassium hydroxide for anions and methanesulfonic acid for cations, carries the sample. Many modern systems generate eluent electrolytically from water.
  2. Injection. A small sample volume enters the eluent stream, often from an autosampler.
  3. Separation. Guard and analytical ion-exchange columns retain ions by charge and size, so different ions emerge at different times.
  4. Suppression. The suppressor exchanges eluent counterions, converting the eluent to a weakly conducting form and the analytes to a more conductive form.
  5. Detection. A conductivity cell, or another detector, measures each ion as it elutes, producing a chromatogram.

A suppressor patent summarizes the suppressor’s purpose: to reduce the conductivity and noise of the background while enhancing the conductivity of the analytes, increasing the signal-to-noise ratio and maintaining chromatographic efficiency. Early suppressors were packed columns needing frequent offline regeneration; most modern systems use continuously, electrolytically regenerated suppressors. Suppressor and eluent generator upkeep are covered in IC System Maintenance: Suppressors, Eluent Generators, and Column Life.

Interested in chemistry and materials science?

Register for a FREE Lab Manager account to subscribe to our Chemistry & Materials Science Newsletter.
Subscribe for Free

Suppressed or non-suppressed conductivity?


Suppresssed conductivityNon-supressed Conductivity

How it works

A suppressor reduces eluent background before detection

Low-capacity columns and low-conductivity eluents allow direct detection

Sensitivity

Higher; lower background noise

Lower, but adequate for many applications

Complexity

More components to maintain

Simpler hardware

Typical use

Trace anions, regulatory methods

Routine cation and higher-concentration analyses

 

Non-suppressed detection followed soon after the original method. As a general history notes, a non-suppressed method for anions was introduced in 1979 by Gjerde and colleagues, followed by cations in 1980, setting off a period of competition between the two approaches. Both remain in use today.

Which detectors are used?

DetectorTypical Use

Conductivity (suppressed or non-suppressed)

Most inorganic anions and cations

Amperometric (including pulsed amperometric)

Carbohydrates, amines, and other electroactive species

UV-visible, direct or after post-column reaction

Nitrite, bromate, transition metals, and other UV-active or derivatized species

Mass spectrometry (IC-MS)

Confirmation and identification of ionic species

ICP-MS (IC-ICP-MS)

Elemental speciation, such as arsenic or chromium species

 

Ion chromatography or ICP?

They answer different questions. ICP-OES and ICP-MS measure total elements after digestion; see our guide to microwave digestion systems for that workflow. IC measures specific ions and species. It can tell nitrate from nitrite, bromide from bromate, and, as the cation review notes, provide metal speciation information that spectroscopic techniques alone can’t. Many labs use both, and IC-ICP-MS combines them for speciation work. The NBS review identified speciation by oxidation state and complexation as an important goal for the field.

IC is also often confused with ion exchange chromatography used for protein purification. They share a separation principle but serve very different purposes; see Ion Chromatography vs. Ion Exchange Chromatography: Clearing Up the Confusion.

What to look for in an IC system

FeatureWhy it Matters

Anion, cation, or dual channel

Dual systems run anions and cations simultaneously

Eluent generation (reagent-free IC)

Produces consistent eluent from water, reducing preparation errors and labor

Suppressor type

Electrolytic, chemical, or none; affects sensitivity and maintenance

Detector options

Conductivity plus amperometric, UV, or MS for broader applications

Inline sample preparation

Filtration, dialysis, or dilution handles dirty or high-matrix samples

Column formats

Standard, microbore, or capillary columns trade sensitivity, speed, and eluent use

Autosampler and throughput

High-volume labs need unattended operation

Software and compliance

Audit trails and validated software for regulated work

Service and consumables

Columns, suppressors, eluent cartridges, and support

 

Water quality matters as much as hardware. IC eluents and blanks depend on ultrapure water, so monitor your water system; our TOC analyzer buyer’s guide covers organic carbon specs for Type I water. Reagent purity logic is similar to chromatography generally; see our guide to choosing solvent grades.

An ion chromatography readiness checklist

  1. Target ions and methods are defined. List the anions, cations, and species you must measure, and the regulatory or compendial methods that apply.
  2. Concentration ranges are known. Trace work in ultrapure water and percent-level analyses in process samples need different columns, loops, and detection.
  3. Suppressed vs. non-suppressed is decided. Choose suppressed conductivity for trace and regulatory anion work; non-suppressed may suit simpler or higher-concentration applications.
  4. Detector needs extend beyond conductivity where required. Carbohydrates need amperometric detection; some species need UV or MS; speciation may call for ICP-MS coupling.
  5. Sample matrices are characterized. High-matrix or particulate samples need filtration, dialysis, or dilution, ideally inline.
  6. Water and reagent quality are controlled. Ultrapure water and clean eluents keep baselines stable and blanks low.
  7. Throughput and automation fit the workload. Dual-channel systems and autosamplers suit high-volume labs.
  8. Maintenance and consumables are budgeted. Columns, suppressors, and eluent cartridges have finite lives; plan their replacement.

Ion chromatography rarely gets the attention of mass spectrometry, yet it quietly measures the ions that matter most in water, food, and medicines. Match the system to your ions, matrices, and methods, keep the water clean, and it will remain one of the most dependable tools in the lab.

 

This article was produced under Lab Manager’s AI Editorial Guidelines

Frequently Asked Questions (FAQs)

  • What is ion chromatography used for?

    Ion chromatography measures inorganic anions, cations, and small ionic organic species. Common uses include drinking water and wastewater anions, cations such as ammonium and sodium, pharmaceutical counterions, food ions and carbohydrates, and trace ions in ultrapure water for power and semiconductor plants.

  • How does ion chromatography work?

    A sample is carried by an ionic eluent through ion-exchange columns that separate ions by charge and size. A suppressor then lowers the eluent’s background conductivity and enhances the analyte signal before a conductivity detector measures each ion.

  • What is a suppressor in ion chromatography?

    A device after the separation column that converts the eluent to a weakly conducting form and the analytes to a more conductive form, increasing the signal-to-noise ratio for conductivity detection.

  • What is the difference between ion chromatography and ICP?

    ICP-OES and ICP-MS measure total element concentrations. Ion chromatography measures specific ionic species, such as nitrate versus nitrite or bromide versus bromate, and can provide speciation information.

  • Who invented ion chromatography?

    Modern ion chromatography with suppressed conductivity detection was introduced in 1975 by Hamish Small, T. S. Stevens, and W. C. Bauman of Dow Chemical.

Add Lab Manager as a preferred source on Google

Add Lab Manager as a preferred Google source to see more of our trusted coverage.

About the Author

  • Trevor Henderson headshot

    Trevor Henderson BSc (HK), MSc, PhD (c), has more than two decades of experience in the fields of scientific and technical writing, editing, and creative content creation. With academic training in the areas of human biology, physical anthropology, and community health, he has a broad skill set of both laboratory and analytical skills. Since 2013, he has been working with LabX Media Group developing content solutions that engage and inform scientists and laboratorians. He can be reached at thenderson@labmanager.com.