Key takeaways
- ICH Q3D sets permitted daily exposures (PDEs) for elemental impurities and requires a risk assessment; USP <232> sets the limits and USP <233> sets the analytical procedures.
- USP <233> allows four forms of sample preparation: neat, direct aqueous solution, direct organic solution, and indirect solution, with total digestion preferred for indirect solutions.
- Closed-vessel digestion minimizes losses of volatile impurities such as mercury and is the usual route for insoluble samples.
- Quantitative procedures are validated with spikes added before sample prep at 50% to 150% of the target concentration (J), with mean recoveries of 70% to 150% and repeatability of 20% RSD or less.
- Risk assessments commonly use a control threshold of 30% of the PDE to decide whether routine testing is needed.
For decades, pharmaceutical heavy metals testing under USP <231> relied on a colorimetric comparison. It was simple, but it couldn’t tell one element from another and was prone to losses. As Agilent’s white paper on the transition notes, the updated USP chapters <232> and <233>, aligned with ICH Q3D, were implemented in January 2018 in place of that colorimetric test. In their place came element-specific instrumental methods, permitted daily exposure limits, and a much closer look at how samples are prepared.
For ICH Q3D and USP <232>/<233> testing, sample preparation means choosing among four forms set out in USP <233>, neat, direct aqueous solution, direct organic solution, or indirect solution, with closed-vessel total digestion the preferred route for materials that won’t dissolve. The harmonized USP <233> text official May 1, 2026 states that total digestion is the preferred approach to obtain an indirect solution and that closed vessel digestion minimizes the loss of volatile impurities. This article explains how the risk assessment shapes testing, how to choose a preparation, and what validation requires. For digestion systems, see our guide to microwave digestion systems.
How does ICH Q3D shape what you test?
ICH Q3D starts with toxicology and risk, not testing. It sets PDEs by route of administration and groups elements into classes. As one reference materials supplier summarizes, Class 1 includes arsenic, cadmium, mercury, and lead, and Class 2A includes cobalt, nickel, and vanadium, with further classes for elements that matter mainly when intentionally added or for particular routes. The current version, ICH Q3D(R2), finalized in April 2022, corrected some PDEs and added limits for cutaneous and transcutaneous products.
- Identify potential sources: APIs, excipients, catalysts, water, manufacturing equipment, and container closure systems.
- Estimate levels from supplier data, published data, or testing.
- Compare with the control threshold. A risk assessment guide notes that a control threshold of 30% of the PDE is commonly used; below it, additional controls may not be needed.
- Test where needed, using USP <233> or an equivalent validated procedure.
Methods must be sensitive enough to support that threshold. A Shimadzu application note on cutaneous products under Q3D(R2) demonstrates measuring Class 1 and 2A elements well below 30% of the PDE.
What sample preparation does USP <233> allow?
| Form | When it's used | Notes |
|---|---|---|
Neat | Liquids analyzed without dilution | Rare for drug products; viscosity and matrix effects limit use |
Direct aqueous solution | Sample dissolves in an aqueous solvent | Simple; check that all target elements stay in solution |
Direct organic solution | Sample dissolves in an organic solvent | Requires instrument setup suited to organic matrices |
Indirect solution | Material isn’t directly soluble in aqueous or organic solvents | Total digestion, typically closed-vessel, is preferred |
The USP <233> chapter text describes the closed vessel digestion procedure: a weighed sample in concentrated acid, allowed to sit loosely covered in a fume hood, then digested in a closed vessel until digestion or extraction is complete, following the manufacturer’s recommended procedures for safe operation. The harmonized text adds that the choice of acid depends on the sample matrix.
Why closed-vessel digestion?
- Volatile elements stay in the vessel. Mercury, and in some cases osmium and other volatile species, can be lost from open digestions.
- Insoluble matrices dissolve completely, including many excipients and coated tablets.
- Contamination is lower because vessels are sealed and acid volumes are small.
- Larger sample masses can be digested with little dilution, which helps meet sensitivity targets; Agilent notes that closed-vessel digestion in nitric acid enables this while stabilizing regulated elements in solution.
- Methods are well established. Application notes show validated closed-vessel workflows for products such as antibiotics.
What is J, and why does it matter for sample prep?
J is the target concentration of an element in the prepared sample solution, at the limit. It’s derived from the element’s PDE, the product’s maximum daily dose, and the dilution used in sample preparation. Every preparation choice, sample mass, final volume, and dilution, changes J. Too much dilution can push J below what the instrument can reliably measure; too little can overload the plasma with matrix. Plan sample mass and dilution against J before validation.
What does USP <233> validation require?
| Parameter | Requirement for quantitative procedure |
|---|---|
Accuracy | Spike the material with reference materials before any sample preparation steps, at 50% to 150% of J; mean recovery of three preparations at each level must be 70% to 150% |
Repeatability | Six independent preparations spiked at the target; relative standard deviation no more than 20% |
Ruggedness or intermediate precision | Repeat on a different day, instrument, or analyst, per the chapter |
Specificity | Demonstrate the method distinguishes each target element from others and from the matrix |
The requirements are from the USP <233> chapter text and harmonized revision. A Thermo Fisher presentation on the chapters outlines the validation sequence, including limit procedures that check detectability with spiked solutions. The key prep point is when spiking happens: spikes go in before digestion or solubilization, so the validation tests the whole procedure, not just the instrument.
Practical sample prep challenges
- Mercury stability. Mercury can be lost or adsorb to surfaces; stabilizing it in solution, often with hydrochloric acid, is common practice.
- Platinum-group catalysts. Palladium, platinum, and related elements may need hydrochloric acid or aqua regia to stay in solution.
- Silicates and titanium dioxide. Some excipients resist nitric acid and may need hydrofluoric acid, with its added hazards; see Microwave Digestion Safety: Pressure, Acid Handling, and Vessel Venting.
- Blanks. At the low levels required, reagent and labware contamination can dominate results; see Controlling Blanks in Trace Metals Sample Prep for ICP-MS.
- Sample mass limits. Organic-rich samples generate gas; respect vessel limits for sample mass.
- Matrix effects. High dissolved solids can suppress signals; balance digestion mass against dilution.
An elemental impurities sample prep checklist
- The Q3D risk assessment defines which elements need testing. Testing every element by default wastes effort. Use the assessment and control threshold to target elements and products.
- J is calculated before method development. J depends on PDE, maximum daily dose, and dilution. Calculate it first so sample mass and dilution are chosen to meet sensitivity needs.
- The preparation form fits the material. Use direct solutions when samples dissolve completely; use closed-vessel total digestion for insoluble materials.
- Acids suit the elements and matrix. Nitric acid is the base; add hydrochloric acid for mercury and platinum-group elements, and HF only when needed for silicates, with proper controls.
- Spikes are added before sample preparation. Validation must cover the whole procedure, including digestion.
- Validation meets USP <233> criteria. Mean recoveries of 70–150% at 0.5–1.5J, repeatability RSD of 20% or less for six preparations, plus ruggedness and specificity.
- Blanks and contamination are controlled. Ultrapure acids, clean labware, and method blanks protect low-level results.
- Current harmonized text is used. USP <233> was harmonized with an official date of May 1, 2026. Check that SOPs reflect the current text.
Elemental impurities testing moved from a color comparison to a risk-based, element-specific system, and sample preparation is where that system succeeds or fails. Let the risk assessment set the targets, calculate J first, digest insoluble materials completely in closed vessels, and validate the whole procedure, from spike to signal.