Ultra-Low Freezers: Energy, Capacity, and the Case for a –70 °C Setpoint

Many labs once stored samples at −70 °C. The −80 °C default crept in later, and a researcher who spent years searching for evidence that −80 °C preserves samples better found scant support for a difference.

Written byTrevor J Henderson
| 5 min read
Green labs coordinator adjusting an ultra-low freezer control panel with a plug-in electricity meter on top of the unit.
Register for free to listen to this article
Listen with Speechify
0:00
5:00

Key takeaways

  • An ultra-low temperature (ULT) freezer typically stores samples at −70 to −80 °C, and raising a suitable unit from −80 °C to −70 °C commonly cuts its energy use by about 20% to 40%.
  • Researchers have found little evidence that −80 °C preserves most samples better than −70 °C, and many institutions now run large shares of their ULT fleets at −70 °C.
  • A −70 °C setpoint leaves less margin before alarm or transfer thresholds during a failure, so it should come with strong monitoring and a contingency plan.
  • Capacity management, including clearing out old samples and consolidating partly full units, often saves more energy and space than buying new freezers.
  • ULT freezers use cascade, auto-cascade, or Stirling cooling, and many newer models use natural refrigerants; ENERGY STAR ratings help compare efficiency.

 

At the University of Colorado Boulder, labs that had been storing sensitive samples at −70 °C for years started asking a simple question: why does everyone else use −80 °C? As a 2018 feature in ACS Central Science recounts, one researcher who spent years searching the peer-reviewed literature for evidence that −80 °C preserves samples better than −70 °C found scant evidence for a difference. The colder setting had spread largely as a default, not as a scientific requirement.

That matters because an ultra low freezer is among the most energy-hungry pieces of lab equipment. For most samples, a −70 °C setpoint protects them as well as −80 °C while cutting energy use substantially, though the change brings a smaller safety margin that needs managing. This article covers how ULTs work, what they cost to run, the evidence and trade-offs behind −70 °C, and how to get more from existing capacity. For the wider landscape, see our comparison of lab cold storage options.

How do ultra-low freezers work?

Technology

How it works

Typical notes

Cascade refrigeration

Two compressors and refrigerant circuits work in series, the first pre-cooling the second

Common in upright ULTs; robust recovery

Auto-cascade refrigeration

A single compressor with a refrigerant blend that separates internally

Simpler design, often in smaller units

Stirling engine

A free-piston engine with a sealed helium working fluid

No conventional compressors; often energy efficient

 

These categories follow one supplier’s overview of ULT designs. Many newer ULTs use natural hydrocarbon refrigerants with low global warming potential, and insulation improvements such as vacuum insulated panels increase usable space for the same footprint.

How much energy does an ultra-low freezer use?

The U.S. Department of Energy’s Federal Energy Management Program estimates that a conventional ULT uses about 20 kWh per day, and its purchasing guidance puts annual use at roughly 1,916 kWh for the best available models, 5,019 kWh for ENERGY STAR-qualified models, and 7,300 kWh for less efficient ones. Freezers also release that energy as heat, adding cooling load to the room and building.

Does −70 °C protect samples as well as −80 °C?

For most common sample types, the available evidence suggests yes. The ACS Central Science feature describes long-term −70 °C storage of sensitive samples at CU Boulder, and notes that the University of Edinburgh began a five-to-10-year study comparing −80, −70, and −60 °C storage. CU Boulder’s Green Labs program maintains a database, built with contributors from UC Davis and other campuses, documenting sample types stored at −70 °C, and reports that CDC raised 60 ULT freezers on one campus to −70 °C as part of a freezer challenge.

There are exceptions. Some protocols, regulatory filings, sponsors, or product labels specify −80 °C or a particular range, and those requirements take precedence. Check sample requirements, especially for regulated or long-term archival material, before changing setpoints.

How much energy does a −70 °C setpoint save?

Source

Reported effect of moving from −80 °C to −70 °C

University of Copenhagen test

Total plug load 20% to 22% lower; cabinet temperature also more stable at −70 °C

E Source summary of field test data

High-efficiency models averaged 0.39 kWh/ft³/day at −80 °C and 0.20 at −70 °C

ACS Central Science feature

Tests by researchers and My Green Lab suggest up to 40% less energy

 

Sources: the University of Copenhagen test report, E Source’s summary, and the ACS Central Science feature. Savings vary by freezer age, design, condition, and loading, so meter your own units where possible. Adoption has grown accordingly: by the time of the 2018 feature, roughly half of CU Boulder’s approximately 160 ULT freezers ran at −70 °C, compared with 3% in 2010.

Weighing the trade-offs

The main concern is time. A freezer at −70 °C starts 10 °C closer to alarm and transfer thresholds if it fails. CU Boulder notes manufacturer data suggesting an empty ULT warms to −50 °C about one to one and a half hours sooner from a −70 °C setpoint than from −80 °C. On the other side, CU Boulder also points out that compressors working less hard may face less risk of failure. The practical response is to pair a −70 °C setpoint with reliable remote alarms, set appropriate alarm thresholds, and a tested plan for moving samples; see  Freezer Temperature Monitoring and Alarms: Meeting GMP, CAP, and Grant Requirements and  When the –80 Goes Down: A Freezer Failure Contingency Plan.

Getting more from the capacity you have

  • Clear out old samples. Abandoned and duplicate samples fill many freezers. Programs such as the International Laboratory Freezer Challenge encourage clean-outs, inventory, and setpoint changes.
  • Consolidate partly full units. Combining contents can free whole freezers to be retired or kept as backup capacity.
  • Use racks and inventory systems. Standardized racks and a sample inventory reduce door-open time and make full use of shelves.
  • Defrost and maintain. Ice buildup on gaskets and inner doors reduces efficiency and can stop doors sealing; clean condenser filters regularly.
  • Manage room conditions. Freezers work harder in hot, crowded rooms. Leave clearance for airflow and plan room cooling for clusters of units.
  • Keep one unit as backup. Empty or partly empty backup capacity is the cheapest insurance in a failure.

Making the switch to −70 °C

  1. Confirm that stored samples, protocols, and any regulatory or sponsor requirements allow −70 °C.
  2. Meter energy use before and after the change on a representative unit, if possible.
  3. Adjust alarm thresholds and confirm remote alerting works.
  4. Update the freezer failure plan to reflect the shorter margin.
  5. Label freezers with their setpoint and communicate the change to all users.
  6. Review the setting periodically, and document it for regulated storage, where qualification and mapping may need updating; see Lab Manager’s guide to GMP-compliant cold storage.

An ultra-low freezer checklist

  1. Sample requirements are documented for each freezer. Knowing what each unit holds, and at what temperature it must be stored, is the basis for setpoint and capacity decisions. A simple inventory by freezer is enough to start.
  2. Setpoints are chosen deliberately. Use −70 °C where samples allow, and −80 °C where requirements specify it. Avoid defaulting to the coldest setting out of habit.
  3. Energy use is known. Metering, or ENERGY STAR data, shows which units cost most to run. Older, inefficient units are often the best candidates for replacement or retirement.
  4. Monitoring and alarms match the setpoint. Independent loggers, remote alerts, and thresholds set with the setpoint in mind give staff time to act. This matters more at −70 °C.
  5. Backup capacity exists. An empty or partly empty unit, or shared backup across groups, shortens response time during a failure.
  6. Clean-outs and consolidation happen regularly. Scheduled clean-outs keep capacity available and can free entire freezers for retirement or backup.
  7. Maintenance is scheduled. Defrosting, gasket checks, and condenser filter cleaning keep units efficient and reliable.
  8. New purchases are compared on efficiency. ENERGY STAR certification, cooling technology, and usable capacity per kWh should be part of every purchase decision.

Ultra-low freezers protect some of a lab’s most valuable material, and they’ll keep drawing power for a decade or more. Store at the temperature samples really need, fill the space you have before buying more, and back the setpoint with monitoring that buys time when something goes wrong.

 

This article was produced under Lab Manager’s AI Editorial Guidelines 

Frequently Asked Questions (FAQs)

  • Is −70 °C cold enough for long-term sample storage?

    For many common sample types, including DNA, RNA, proteins, and cells in appropriate media, available evidence suggests −70 °C performs comparably to −80 °C. Follow any protocol, regulatory, or sponsor requirement that specifies −80 °C.

  • How much energy does a −80 °C freezer use per day?

    A conventional ultra-low freezer uses about 20 kWh per day, according to U.S. Department of Energy guidance. Efficient and ENERGY STAR-qualified models use considerably less.

  • What refrigerant is used in ultra-low freezers?

    It depends on the design. Cascade and auto-cascade freezers use refrigerant circuits, and many newer models use natural hydrocarbon refrigerants. Stirling engine freezers use a sealed helium working fluid instead of conventional refrigerants.

  • Does a −70 °C setpoint shorten the time to react if a freezer fails?

    Yes. A freezer starting at −70 °C reaches warning thresholds sooner than one starting at −80 °C, so pair the setpoint with remote alarms and a tested contingency plan.

  • What is the International Laboratory Freezer Challenge?

    It’s a program that encourages labs to improve cold storage management through practices such as clean-outs, inventory, maintenance, and raising ULT setpoints to −70 °C where appropriate.

Add Lab Manager as a preferred source on Google

Add Lab Manager as a preferred Google source to see more of our trusted coverage.

About the Author

  • Trevor Henderson headshot

    Trevor Henderson BSc (HK), MSc, PhD (c), has more than two decades of experience in the fields of scientific and technical writing, editing, and creative content creation. With academic training in the areas of human biology, physical anthropology, and community health, he has a broad skill set of both laboratory and analytical skills. Since 2013, he has been working with LabX Media Group developing content solutions that engage and inform scientists and laboratorians. He can be reached at thenderson@labmanager.com.

    View Full Profile