What Your System Suitability Test Is Telling You Before It Fails

How lab leaders can turn routine system suitability data into early warnings for column failure, batch disruption, and data integrity risk

Written bySarah BauderPresented byNazmul Alam
| 5 min read
Laboratory professional performing maintenance on analytical equipment
Register for free to listen to this article
Listen with Speechify
0:00
5:00

It is 8:15 on a Monday morning. The system suitability test has just failed, 90 samples are waiting, and the record shows nothing but passes. Yet the tailing factor had climbed from 1.1 on a new column to 1.6 a month earlier. Because both under the 2.0 limit, both are filed as passes. "And the numbers were just sitting there," said Nazmul Alam, PhD, an analytical chemistry consultant and trainer with 14 years of HPLC and LC-MS experience. Speaking at the 2026 Mass Spectrometry & Advanced Separations Digital Summit, Alam delivered a message for lab leaders: trending the suitability data you already collect is one of the cheapest ways to turn surprise column failures into scheduled maintenance—and to protect data integrity.

The danger of the pass/fail mindset

Most labs file system suitability testing (SST) as a simple pass or fail. Alam demonstrated what happens when the same results are plotted instead. In his illustrative schematic, the tailing factor (which USP now calls the symmetry factor) holds near 1.1 while the column is healthy, then climbs for months. Every data point still passes. However, an internal action limit set at 1.5 would have flagged the drift 32 runs, or about six working weeks, before the test failed.

Timing dictates the cost. A failure caught before a batch starts requires a routine maintenance job: change the column, requalify, and run late. A failure caught mid-batch triggers an entirely different protocol. Under the FDA's guidance on out-of-specification (OOS) results, data from the suspect period should not be used. Every injection since the last passing check is drawn into question, and a GMP lab may face a formal investigation.

"It is the same column, the same failure, only the timing is different," Alam said. "So the timing is the part that you can change."

Regulatory backing: The gate vs. the camera

ICH Q14, adopted in November 2023, states the system suitability test confirms a measurement system is fit for its intended purpose, and should also increase the detectability of unacceptable performance. Alam compared the traditional SST approach to a parking lot gate with a camera beside it. "Most laboratories run the suitability test as a gate," he said. "The same test can also be a camera because the record is already there."

Section 6 of Q14 goes further, recommending ongoing monitoring of procedure outputs to identify trends. Alam was careful not to claim that labs are out of compliance if they don't trend. Still, "if somebody asks you why you trend them, the answer is now written in a guideline," he said. "It's easier to defend trending than to defend not trending."

The idea isn't new. John Dolan urged LCGC readers to plot suitability values on control charts back in 2011. "None of this is actually new," Alam noted. "And I think that's the uncomfortable part."

Reading the early warning signs

Not every signal arrives at once. In a second illustrative example covering one column over four months, the tailing factor crossed its action limit at run 57, retention time at run 69, and starting pressure at run 79. Tailing gave 22 more runs of notice than pressure, or more than four working weeks. Labs that watch only pressure are relying on the slowest signal.

Lab manager academy logo

Advanced Lab Management Certificate

The Advanced Lab Management certificate is more than training—it’s a professional advantage.

Gain critical skills and IACET-approved CEUs that make a measurable difference.

Not all SST parameters degrade at the same rate. In a four-month column tracking example, the tailing factor crossed its action limit at run 57, retention time at run 69, and starting pressure at run 79. Tailing provided more than four working weeks of notice compared to pressure. Labs that only monitor pressure are relying on the slowest signal. 

Alam offered a quick reference for what drifting suitability parameters usually indicate:

  • Retention time drifting in one direction: Mobile phase preparation, temperature variations, or an aging column.

  • Tailing factor rising over weeks: Column aging as silanols become active, or a mobile phase pH sitting too close to the analyte's pKa.

  • Plate count falling, often as tailing rises: A void forming at the head of the column, or a column reaching the end of its life.

  • Widening percent RSD across replicate injections: The injector or injection process. As Alam stated, this is "not chemistry at all."

How can labs troubleshoot a drifting system suitability parameter?

When a value moves, most labs start troubleshooting with the instrument. Alam reverses that order with "four checks before the instrument": the question, the vial, the chemistry, and only then the instrument. For a validated method, the suitability test has already answered the question, so start at the vial.

Drawing on a 2016 case Dolan published, Alam recommends running 10 injections from one vial and 10 from a second. If precision is significantly better within one vial, the issue is the vial, not the injector. Retention that shifts with each new mobile phase batch points to preparation. To confirm a hardware issue, put a trusted column on the same instrument. "What moves with the column is chemistry," Alam said. "And what stays with the instrument is hardware."

Interested in lab tools and techniques?

Register for a FREE Lab Manager account to subscribe to our Lab Tools & Techniques Newsletter.
Subscribe for Free

Crucially, this troubleshooting requires logging the column serial number. "A trend belongs to a column, not to a method," he stressed. Without it, every new column resets the trend, turning a year of valuable data into unusable noise. 

Implementation: the limit inside the limit

Alam's core recommendation is establishing a limit inside the limit. If the acceptance criterion for tailing is not more than 2.0, the lab might decide that 1.5 is the threshold to plan a column change. "The acceptance criteria tells you when the test failed, which is useful, but late," he said. "The internal action limit tells you when to act."

Those two values must remain clearly separated in methods and SOPs. "Crossing an internal limit is not a failure," Alam said. "You do not want an auditor reading them as the same thing."

For timing the change itself, Alam cited Dolan's rule of thumb: replace a column at roughly a 30 percent loss in plate count from new. The column is the inexpensive part, at about a dollar per sample by Dolan's calculations. The real expense is a stopped batch with an investigation behind it.

To address fears that trending creates audit liabilities, Alam noted that the values already exist in reviewed raw data. "The real risk is a trend with no defined response," he said. His solution is to build the response into the lab's SOP from the outset. It should name who reviews the trend, how often, where the internal action limit sits, and what happens when it's crossed.

How can a lab start trending system suitability data without new software?

A minimum viable SST trending program requires nothing new to be run. Alam recommends:

  • One row per suitability run: Date, method, instrument, column, column serial number, retention time, tailing factor, plate count, percent RSD, and starting pressure.

  • A defined approach to replicates: Neither USP nor ICH specifies whether to log the mean or every injection. The method should decide before the run, and the individual injections should stay in the data system.

  • An internal action limit inside each acceptance criterion: Documented separately from the acceptance criteria themselves.

  • A named reviewer: Someone who reads the trend at every preventive maintenance visit and after any part replacement. Alam estimates it takes about 15 minutes per month, per method.

For LC-MS labs, free tools such as Skyline with Panorama AutoQC and the open-source MSstatsQC package can chart these metrics automatically. Even so, Alam advises starting with a spreadsheet for a couple of months to learn what you need before making purchasing decisions.

Yet, software won't keep abreast of issues. Citing LCGC columnist Dwight Stoll, Alam noted that, as far as Stoll knows, no commercial chromatography data system currently flags early warnings from pressure data. "So the gap is real, and somebody in your laboratory has to be the one who is watching."

The ultimate goal: stop testing into compliance 

When asked about the most common SST mistake, Alam didn't hesitate: treating the acceptance criteria as the target. That mindset drives the riskiest response to a morning failure, which is re-injecting until something passes. He warned that this "turns a maintenance signal into a data integrity problem." The FDA's OOS guidance takes the same line, calling repeated testing until a result passes "testing into compliance." Instead, labs must confirm the failure is real, check the history of the failed value, and have an authorized individual make a documented run-or-change decision. 

For lab leaders, the payoff is concrete. By comparing performance against historical records rather than absolute limits, critical decisions no longer fall to an analyst standing in front of a 90-sample queue. 

As per Alam's closing advice: "Compare against your own record, and not against the acceptance criteria. Because a tailing factor of 1.9 against a limit of 2 is [a] passing test and a failing column."

Frequently Asked Questions (FAQs)

  • Why should labs trend system suitability results?

    Trending reveals gradual drift—such as a tailing factor rising as a column ages—weeks before a test fails. That lead time lets labs schedule column changes instead of losing a batch mid-run or opening an out-of-specification investigation.

  • What is an internal action limit in system suitability testing?

    An internal action limit is a lab-defined threshold set inside the acceptance criterion, such as a tailing factor of 1.5 against a limit of 2.0. Crossing it triggers planning, not a failure, so it should be documented separately from the acceptance criteria.

  • What should a lab do when a system suitability test fails?

    Don't keep re-injecting until a run passes. Confirm the failure is real by checking the standard, mobile phase, column, and lines for air. Review the failing parameter's history, then have someone with authority make and document the decision to run or change the column.

  • When should an HPLC column be replaced?

    Alam pointed to John Dolan's rule of thumb: replace a column at roughly a 30 percent loss in plate count compared with when it was new. More broadly, Dolan recommends replacing any part with a predictable lifespan at 70 to 80 percent of its expected lifetime. Both approaches require a baseline recorded when the column was installed, tracked by its serial number.

Add Lab Manager as a preferred source on Google

Add Lab Manager as a preferred Google source to see more of our trusted coverage.

About the Author

  • Placeholder image

    Sarah Bauder is the senior editor at Lab Manager. She possesses a diverse background spanning editorial, digital marketing and film and television production. She brings over 15 years of experience in editorial writing, B2C and B2B content creation. A student of history, she graduated from York University in Toronto, Ontario, Canada. She can be reached at sbauder@labmanger.com.

    View Full Profile

Speaker

  • Headshot

    Nazmul Alam

    Analytical Chemistry Consultant and Trainernalam.ca

Related Topics

Loading Next Article...
Loading Next Article...
Current Magazine Issue Background Image

CURRENT ISSUE - October 2026

Why Lab Teams Resist Change—And How Leaders Can Respond

Addressing the human side of automation, AI, and operational change

October 2026 Lab Manager Cover Image