Laboratories that reach the detection limit of their immunoassays are often steered toward a new ultra-sensitive platform, a capital purchase that is harder to justify as equipment budgets tighten. Beyond the price tag, that decision can bring proprietary consumables, costly revalidation of established workflows, and sensitivity gains that don't always hold up in real sample matrices. In this Q&A, Jonathan Royce shares how lab managers can determine how much immunoassay sensitivity they actually need, evaluate the evidence behind amplification approaches, and weigh the cost of a new platform against improving the workflow they already have.
Many labs are working with reduced equipment budgets this year. When a lab reaches the detection limit of its current immunoassay workflow, how does that conversation usually unfold, and where do lab managers tend to get stuck?
Most researchers assume that if their assay can't quantify low-concentration samples, they either have to redevelop the assay, discard the low-concentration samples, or report them as “less than” a value with no real quantification. Unfortunately, because redevelopment takes time and money, the latter two options are common. Even in peer-reviewed studies, it is not unusual to see unquantifiable samples excluded or results reported below the assay's validated range. The validated workflow is another hurdle. Changing an antibody pair or instrument can mean revalidating years of work, which makes labs reluctant to switch.
For lab managers on tight budgets, the first question is whether a new platform is necessary at all. My advice is to start with the data: identify how many samples fall below the lower limit of quantification (LLOQ), and by how much. A threefold sensitivity improvement requires a very different investment from a hundredfold one. Then compare the options: optimizing the current assay, adding a reagent-level amplification step, or moving to a new platform.
Beyond the purchase price, what should a lab manager add up before committing to a new ultra-sensitive platform? Which costs tend to surprise people a year or two in?
The main consideration should be whether you can accept working with proprietary reagents or want to retain the flexibility to work with your own antibodies, reagents and other assay components. Many instrument-based solutions for ultra-sensitivity are designed to lock users into a closed ecosystem.
Lost flexibility is hard to quantify in dollars, but it can significantly limit the options available for future projects. You should also factor in instrument maintenance and calibration, as well as proprietary reagents and consumables. For example, an ELISA plate costs about $3, while plates for proprietary instruments can cost more than 30 times as much. The surprises tend to arrive later: service contracts once the warranty ends, changes in consumable prices, new targets the vendor doesn't support, collaborators who can't reproduce your results without the same instrument, and equipment that sits idle between projects.
How should a lab work out how much additional sensitivity it actually needs? What is the risk of buying more sensitivity than the application calls for?
Consider where your samples below the assay's LLOQ tend to fall, and add some margin. You want to be in the linear part of the assay curve, which ensures good recoveries with low variability. At the same time, running an assay that is “too sensitive” may leave some samples above the upper limit of quantification. Every assay has a dynamic range, which normally covers concentration ranges between 1000-100,000x (i.e., 3 to 5 logs). Ultra-sensitive technologies may shift the lower limit down, but they will also bring the upper limit down by a similar degree.
Some approaches raise assay sensitivity without new instrumentation. What evidence should a lab manager ask for before trusting one in an established workflow?
It is important to verify that the chosen approach produces repeatable results and does not increase the variability between repeated measurements of the same sample. Additionally, the approach needs to be compatible with the sample matrix, be it whole blood, serum, plasma or cerebrospinal fluid (CSF). Some approaches look great in buffer, but real samples come in complex matrices that can produce unacceptably high background. In practice, ask how the detection and quantification limits were calculated, and whether the comparison was matched, using the same antibodies and samples. Then ask for spike recovery, dilution linearity, and intra- and inter-assay coefficients of variation (CVs), all generated in the matrix your lab actually uses.
A matched comparison was published in SLAS Technology this spring. What did the study show, and what questions does it leave open for labs?
The study describes the development of Binding Oligo Ladder Detection, or BOLD, which is our molecular amplification technology for standard immunoassays. We used an ELISA for SARS-CoV-2 spike protein as a model system and applied BOLD, which increased the assay's sensitivity 67-fold. While this is a substantial improvement, the gain will vary by assay and application.
The main question the paper leaves open is how broadly applicable BOLD is across different assays. We have now applied BOLD to around 30 biomarkers spanning neurology, cardiology, oncology, infectious disease and inflammation, with results normally ranging from 10-100x improvement in sensitivity. In future work, we would like to apply BOLD in a full validation study to further demonstrate its robustness.
How do sample type and volume change the decision? What happens to sensitivity gains when a lab moves from buffer to plasma or serum?
Both play a role. The sample type, or matrix, can affect background, choice of antibodies, and which blocking agents and diluents are used in the assay. The matrix will also affect which options are practical for quantifying low-concentration samples.
Sensitivity gains measured in buffer may not translate directly to plasma or serum. Matrix effects can increase background and require sample dilution, reducing analyte concentrations before the assay even starts. That's why the key figure to ask for is the sensitivity improvement demonstrated in the matrix your lab actually uses.
Sample volume is equally important. We tend to think of sample donors as adult humans, but that is often not the case. Children and even infants can be donors, and the volume that can be safely drawn from smaller humans is limited. As a result, samples are often diluted with buffer, which provides enough volume for reliable pipetting but also reduces the concentration of all analytes proportionally. This effect is even more pronounced with small mammals such as mice, rats or rabbits.
Many labs have years of data tied to established workflows. How should lab managers weigh the value of improved sensitivity against the cost and complexity of changing a validated workflow, including comparability and revalidation?
It is not trivial to implement a new platform, especially if you are forced to move away from previously validated antibody pairs. Method validation and instrument validation are costly endeavors, and there should always be an economic analysis underpinning the decision to invest in a change. This is especially true now that options exist that allow labs to keep using their validated antibody pairs and instruments while still improving their lower limits of quantification. Reagent-based solutions still require revalidation, since any change to an assay should be validated to ensure robustness. But the fundamental building blocks of an assay are the ligands (e.g., antibodies) and the instruments used to measure signal, and revalidating those components is more resource-heavy.
If a lab manager could keep one rule of thumb for choosing between a new platform, improving sensitivity within the current workflow, or staying with what they have, what would it be?
Aim for good enough. Most studies do not require single-molecule sensitivity, so invest in what your workflows actually need. As with many things in life, the “keep it simple” rule of thumb applies here.