When the –80 Goes Down: A Freezer Failure Contingency Plan

In 2017, a freezer at the University of Alberta failed, and so did the system meant to report it. Nobody knew until a fire alarm went off. By then, 12.8% of Canada’s ice core archive had been damaged.

Written byTrevor J Henderson
| 5 min read
Technician transferring a frosted sample rack from a failed ultra-low freezer into a cooler of dry ice.
Register for free to listen to this article
Listen with Speechify
0:00
5:00

Key takeaways

  • A freezer failure contingency plan defines how failures are detected, who responds, where samples go, how to buy time, and how to recover and document the event.
  • How long a −80 °C freezer stays cold depends heavily on load, door seals, and whether the door stays shut; one NIH-cited estimate is that an average new ULT can warm to −60 °C in about five hours.
  • Monitoring that isn’t tested end to end can fail silently, as it did in the 2017 University of Alberta ice core loss.
  • Dry ice buys time but releases carbon dioxide, so use it with ventilation, quantity limits, and proper handling.
  • Backup freezer capacity, emergency power, split sample storage, and planned replacement of aging units reduce both the chance and the cost of a failure.

 

On a weekend in April 2017, the chillers serving a walk-in freezer at the University of Alberta shut down. The freezer’s computer system sent alarm signals that the temperature was rising, but, as CBC reported, the signals didn’t reach the university’s service provider or its on-campus control centre. The problem surfaced only when firefighters responded to a heat alarm, with the freezer at 40 °C. The university confirmed that 12.8% of the Canadian Ice Core Archive, more than 80,000 years of climate history in 1.4 kilometres of ice, had been damaged.

A freezer failure contingency plan exists to make sure that sequence never plays out. It sets out how a failure will be detected, who responds and how fast, where samples will go, how to buy time while they move, and how to recover and learn afterward. This article walks through each part. For the wider cold storage picture, see our comparison of lab cold storage options.

How long will a −80 °C freezer stay cold without power?

Less time than many people assume, and the answer varies widely. Boston University’s risk management guidance cites the NIH estimate that the average new ULT freezer can warm to −60 °C in only five hours after a failure, as its ULT risk management guide notes. Other reports are more optimistic for ideal conditions: one university facilities notice cited a distributor’s claim that a new, full −80 °C freezer can hold critical temperatures for up to 36 hours if seals are free of frost and the door stays shut, according to UW–Madison’s Russell Labs.

The difference comes down to conditions. A lab safety discussion on freezers and power outages recorded wide variation between units in the same outage, with full freezers and good seals faring much better than units with frosted or worn seals and misadjusted closing mechanisms. What helps:

  • Keep the door closed. Every opening speeds warming.
  • Keep freezers full, using frozen fillers if needed, since thermal mass slows warming.
  • Maintain seals. Frost and damaged gaskets let cold escape quickly.
  • Know your setpoint. Units run at −70 °C have less margin than those at −80 °C; see our guide to ultra-low freezers.

What should a freezer failure plan include?

Element

What it covers

Detection

Independent monitoring, remote alarms, and tested escalation that reach people 24/7

Responders

A call tree with primary and backup contacts, including outside working hours

Inventory

What each freezer holds, priority samples, and where they’re located inside the unit

Backup capacity

Designated empty or partly empty freezers, or shared departmental space

Time-buying measures

Door discipline, dry ice supply, and backup CO2 or LN2 cooling where installed

Power

Which freezers are on emergency generator circuits, and how to request service

Supplies

Coolers, cryogenic gloves, face shields, carts, and labels staged near freezers

Recovery

Sample assessment, documentation, root cause analysis, and insurance reporting

 

Detection deserves special attention. The Alberta loss involved a monitoring system that generated alerts no one received. Test the whole chain, from sensor to the phone in someone’s pocket, on a regular schedule; see our guide to freezer temperature monitoring and alarms.

Lab manager academy logo

Lab Management Certificate

The Lab Management certificate is more than training—it’s a professional advantage.

Gain critical skills and IACET-approved CEUs that make a measurable difference.

What should you do when a −80 goes down?

  1. Confirm the alarm and check the freezer’s actual temperature with an independent reading.
  2. Keep the door closed while you assess, and check power, breakers, and obvious faults.
  3. Call the responder list and facilities or the service provider.
  4. Decide early whether to move samples. If the freezer can’t be restored quickly, move them while temperatures are still low.
  5. Move priority samples first to designated backup space, using pre-chilled coolers and dry ice if needed.
  6. Record times, temperatures, actions, and which samples went where.
  7. Label the failed unit as out of service until it’s repaired and back at temperature.

Speed matters most in the first hours. Plans that pre-assign backup space and stage supplies near freezers let responders act instead of improvise.

Is dry ice a safe backup?

Dry ice can hold frozen samples for a while, but it isn’t hazard-free. As it sublimates, it releases carbon dioxide gas that can build up in small or poorly ventilated rooms. Arizona State University’s cold storage outage guidance advises estimating the dry ice needed before an outage, limiting quantities per 24-hour period, and following dry ice safety instructions. Use insulated gloves and eye protection, never seal dry ice in airtight containers, and consider CO2 monitoring where large amounts are used; see our guide to CO2 and oxygen depletion in the lab.

Backup capacity and backup cooling

  • Designate backup freezers. An empty or partly empty unit, or shared departmental capacity, is the single most useful asset in a failure.
  • Consider backup cooling systems. Some freezers can be fitted with liquid CO2 or LN2 backup systems that inject coolant when temperatures rise, as one manufacturer’s contingency planning white paper describes; these need gas supply checks and maintenance.
  • Put critical freezers on emergency power. Generator-backed circuits are the main protection against power outages, since freezers need hours or days of power that batteries can’t economically supply.
  • Split critical samples. Storing aliquots of irreplaceable material in two freezers, ideally in different rooms or buildings, protects against total loss.

Prevention: maintenance and replacement

Many failures are predictable. Boston University’s guide puts ULT life expectancy at 10 to 15 years and notes that failure rates rise sharply as units wear out, recommending replacement planning before end of life. Regular maintenance, including defrosting, seal checks, and condenser filter cleaning, keeps units reliable in the meantime. The same guide notes that property insurance can’t make up for lost productivity or irreplaceable samples.

After the event

  • Assess sample impact using temperature records and sample-specific stability data.
  • Document the timeline, responses, and decisions.
  • Find the root cause, including whether alarms, call trees, or backups performed as planned.
  • Update the plan, and repair or replace equipment as needed.
  • Report to insurers, sponsors, or regulators where required.

A freezer failure readiness checklist

  1. Alarms are tested end to end. The Alberta loss shows that alarms can trigger without reaching anyone. Test from sensor to responder on a schedule, including after network or staffing changes.
  2. A current call tree covers nights and weekends. Primary and backup contacts, reviewed regularly, make sure someone responds. Shared on-call rotations spread the load.
  3. Each freezer has an inventory with priorities. Knowing what’s inside, and what to move first, saves critical minutes. A simple rack map is enough to start.
  4. Backup capacity is designated in advance. Pre-assigned space avoids a scramble. If space is tight, agree on reciprocal arrangements with neighboring labs.
  5. Emergency supplies are staged near freezers. Coolers, gloves, face shields, and carts nearby let responders act immediately. Check them periodically.
  6. Dry ice sourcing and safety are planned. Know where to get dry ice quickly, how much you need, and how to use it safely in your spaces.
  7. Critical freezers are on emergency power. Confirm which circuits are generator-backed, and label outlets so freezers aren’t moved off them.
  8. Irreplaceable samples are split across units. Duplicate aliquots in separate freezers, rooms, or buildings guard against total loss.
  9. Aging units are scheduled for replacement. Planning replacement before the end of a freezer’s expected life reduces the risk of late-life failures.
  10. The plan is practiced. A short drill, even once a year, reveals gaps in contact lists, supplies, and procedures before a real failure does.

Freezer failures can’t be eliminated, but most losses can. Test the alarms, know where samples will go, keep supplies close, and practice the plan, so that when the −80 goes down, the samples don’t go with it.

 

This article was produced under Lab Manager’s AI Editorial Guidelines

Frequently Asked Questions (FAQs)

  • How long can a −80 °C freezer stay cold without power?

    It varies widely with load, seals, and door discipline. An NIH-cited estimate is that an average new ULT can warm to −60 °C in about five hours, while full units with good seals and closed doors can last considerably longer.

  • What should you do first when a −80 freezer fails?

    Confirm the alarm with an independent reading, keep the door closed while you assess, call responders and facilities, and decide early whether to move samples to designated backup space.

  • Is it safe to use dry ice in a freezer emergency?

    It can be, with precautions. Dry ice releases carbon dioxide as it sublimates, so use it with good ventilation, follow quantity limits, wear insulated gloves and eye protection, and never seal it in airtight containers.

  • How long does an ultra-low freezer last?

    Boston University’s risk management guidance puts the life expectancy of ULT freezers at 10 to 15 years, with failure risk rising as units age.

  • Does insurance cover lost samples after a freezer failure?

    Property insurance may cover equipment, but it can’t replace lost research time or irreplaceable samples, which is why prevention and contingency planning matter.

Add Lab Manager as a preferred source on Google

Add Lab Manager as a preferred Google source to see more of our trusted coverage.

About the Author

  • Trevor Henderson headshot

    Trevor Henderson BSc (HK), MSc, PhD (c), has more than two decades of experience in the fields of scientific and technical writing, editing, and creative content creation. With academic training in the areas of human biology, physical anthropology, and community health, he has a broad skill set of both laboratory and analytical skills. Since 2013, he has been working with LabX Media Group developing content solutions that engage and inform scientists and laboratorians. He can be reached at thenderson@labmanager.com.