Lab Manager | Run Your Lab Like a Business

Whitepaper

Analysis of Selected Anions with HPLC and Electrochemical Detection

Traditionally, conductivity is the standard detection method for ion analysis. During the last 10 years, however, there has been increasing interest in electrochemical detection (ECD) for the analysis of anions.

Traditionally, conductivity is the standard detection method for ion analysis. During the last 10 years, however, there has been increasing interest in electrochemical detection (ECD) for the analysis of anions. This is due to the high sensitivity and selectivity of this type of detection.

Figure 1 shows how certain anions can be selectively detected in the presence of several other anions. In total, 16 anions were injected. Four that can be oxidized: iodide, nitrite, bromide, thiocyanide, and twelve that cannot be oxidized: fluoride, chloride, azide, nitrate, phosphite, sulfate, molybdate, hydrogencarbonate, perchlorite, phosphate, selenate and perchlorate.


In order to demonstrate the selectivity and sensitivity, iodide in table salt was analyzed. There is currently a great deal of interest in the determination of iodide in various foodstuffs, because some data indicates that the consumption of iodide has increased, leading to a potential rise in thyroid disorders. Table salt is almost 100% sodium chloride and with conductivity detection a large chloride peak may overlap with the iodide. Figure 2 shows the selective detection of iodide in a large excess of cloride.

 

Detection potential is of great importance for selectivity and sensitivity and must be determined for each set of analysis parameters.

The Agilent 1049A ECD in combination with an auto-sampler provides a time-saving way to find the optimum potential for several compounds through its autoincrement mode. The potential is automatically increased within a series of runs and the best detection potential can be selected from the chromatogram plots.

An example is given in figure 3. The optimum potential was 1.1 V. At lower potentials the response was insufficient, at higher potentials the detector drift was unacceptable.

The minimum detectable level for iodide was about 40 μg iodide per kilogram table salt. The relative standard deviation for peak heights over 75 runs was 3% for 300 pg of iodide and an injection volume of 10 μl.

Source: Agilent