The Heart of the Matter

Capillary columns have changed the face of GC since their introduction about 35 years ago. The most obvious change involves resolving power: up to 50,000 theoretical plates on a 30-m capillary versus 1500 on a six-foot packed column.

Written byAngelo DePalma, PhD
| 7 min read
Register for free to listen to this article
Listen with Speechify
0:00
7:00

Capillary columns have changed the face of GC since their introduction about 35 years ago. The most obvious change involves resolving power: up to 50,000 theoretical plates on a 30-m capillary versus 1500 on a six-foot packed column.  Capillary columns also permit the rapid heating and cooling that is one hallmark of “fast GC.” Packed columns are still used, however, for high-volume injections and gases in particular or when expert chromatographers wish to experiment with stationary phases.

Major GC system and column manufacturers still emphasize innovation in column technology. “Everyone still pays close attention to the stationary phases, to match column phases to compound classes and reduce bleed,” says Eric Phillips, GC and GC-MS marketing manager at Thermo Fisher Scientific (San Jose, CA). “Column chemistries just keep getting better.”

GC column performance depends on the complex interactions among several factors: column length and diameter, chromatographic conditions like mobile phase flow and temperature, and chemical interactions between analytes and stationary phase.

Selecting a GC column begins with the organic chemistry maxim that like dissolves like. “You should target selectivity first,” advises Chris English, who manages Restek’s (Bellefonte, PA) Innovations Laboratory. “If you’re dealing with glycols, select a phase that’s most like a glycol. If you’re analyzing gasoline, select a nonpolar phase like polydimethylsiloxane.” Stationary phases that dissolve analytes provide the optimal retention, assuming optimization of the remaining conditions.

Obtaining acceptable retention is also possible by using a less-than-optimal stationary
phase in a longer column. Doubling the column length adds approximately 40 percent theoretical improvement in resolution, but the column will cost twice as much and may not last as long as a shorter column.

Similarly, narrower-bore columns resolve analytes more efficiently but require much higher back pressures. Since column capacity is roughly related to the square of column diameter (as with cylinders), loading capacity falls off dramatically from 1 mm to 0.18 or 0.10 mm. This can significantly compress the concentration calibration curve, thus limiting applicable concentration ranges.

To continue reading this article, sign up for FREE to
Lab Manager Logo
Membership is FREE and provides you with instant access to eNewsletters, digital publications, article archives, and more.

About the Author

Related Topics

CURRENT ISSUE - October 2025

Turning Safety Principles Into Daily Practice

Move Beyond Policies to Build a Lab Culture Where Safety is Second Nature

Lab Manager October 2025 Cover Image