Thermal analysis equipment with metallic sample pans

Lab Manager’s Independent Guide to Purchasing a Thermal Analyzer

From polymer melting points to ceramic decomposition: How to choose between TGA, DSC, and DMA.

Written byTrevor J Henderson
| 5 min read
Register for free to listen to this article
Listen with Speechify
0:00
5:00

Executive Summary

"Thermal Analysis" is not a single technique; it is a toolbox. Asking for a "Thermal Analyzer" without specifying the property you need to measure is like asking for a "Vehicle" without specifying if you need to haul cargo or race on a track.

The market is dominated by three distinct acronyms that are often confused. DSC (Differential Scanning Calorimetry) measures heat flow to find melting points and glass transitions. TGA (Thermogravimetric Analysis) measures weight change to find decomposition temperatures and filler content. DMA (Dynamic Mechanical Analysis) measures stiffness to predict how a material handles load at high temperatures.

Purchasing a TGA to find a glass transition temperature (Tg) is a physical impossibility. Purchasing a DSC to measure moisture content is using a sledgehammer to crack a nut.

This guide outlines the physics of heat, mass, and mechanics to ensure you select the right instrument for your material's thermal lifecycle.

1. Understanding the Technology Landscape

Thermal analysis instruments look remarkably similar—often identical boxes with a furnace and an autosampler—but their sensors measure fundamentally different physical properties.

Core Instrument Types

  • DSC (Differential Scanning Calorimeter)The most common thermal instrument. It measures the difference in heat flow between a sample and a reference.
    • Primary Function: Phase Transitions. Melting point, Crystallization, Curing, and Glass Transition (Tg).
    • Best for: Polymers, Pharma, Food, and Quality Control.
  • TGA (Thermogravimetric Analyzer)Essentially a high-precision microbalance inside a furnace. It continuously weighs the sample as it is heated.
    • Primary Function: Composition & Stability. Decomposition temperature, Moisture content, Ash/Filler content.
    • Best for: Rubber, Composites, Ceramics, and stability testing.
  • STA (Simultaneous Thermal Analyzer): A hybrid that combines TGA and DSC into one unit. It measures weight change and heat flow simultaneously on the same sample.
    • Pros: Two answers in one run. Eliminates sample variability.
    • Cons: Expensive. Often has lower sensitivity than a dedicated DSC.
  • DMA (Dynamic Mechanical Analyzer): Clamps a solid sample and bends/twists it while heating.
    • Primary Function: Mechanical Modulus. Viscoelastic properties, Damping (Tan Delta), and Softening points.
    • Best for: Engineering plastics, Rubbers, and structural materials.

2. Critical Evaluation Criteria: The Decision Matrix

The choice of instrument is dictated entirely by the material property that is critical to your product's performance. Use this decision matrix to map your failure mode to the correct hardware.

Decision Track 1: The Analytical Goal

  • "At what temperature does it melt or cure?"DSC
    • Context: You need to know the processing temperature or the degree of cure of an epoxy.
    • Hardware: Heat Flux DSC with a Chiller.
    • Estimated Cost: $35,000 – $60,000
  • "How much glass filler is in this plastic?"TGA
    • Context: You need to burn off the polymer and weigh the ash residue.
    • Hardware: TGA with robust furnace.
    • Estimated Cost: $30,000 – $50,000
  • "Will this part get soft and warp in the summer heat?"DMA
    • Context: You need to measure the modulus (stiffness) drop-off as temperature rises.
    • Hardware: DMA with tension/flexure clamps.
    • Estimated Cost: $50,000 – $80,000

Decision Track 2: Furnace Cooling

Thermal analysis is not just about heating; it's about controlled cooling.

  • Mechanical Chiller (Refrigerated):
    • Range: Down to -40°C or -90°C.
    • Pros: Convenient, no consumables. Cons: Higher initial capital cost.
  • Liquid Nitrogen (LN2):
    • Range: Down to -150°C or -180°C.
    • Pros: Essential for seeing Tg of rubbers/elastomers. Cons: High ongoing operational cost and safety hazards.

3. Key Evaluation Pillars

Once the technique is selected, the "specs" determine the data quality. A baseline that drifts or a balance that jitters will hide the subtle transitions you are looking for.

A. Baseline Stability (DSC)

In DSC, the signal is a flat line until a transition occurs. If the baseline curves or drifts due to furnace design, it can hide weak transitions like a glass transition (Tg).

  • The Test: Ask to see an "Empty Pan" run. It should be flat and straight.
  • Modulated DSC (MDSC): A premium feature that applies a sine wave to the heating rate. It separates "Reversing" (Melting/Tg) from "Non-Reversing" (Curing/Relaxation) events, clarifying complex data.

B. Balance Sensitivity (TGA)

In TGA, the balance is everything.

  • Vertical vs. Horizontal:
    • Vertical: Gravity assists. Often more sensitive but susceptible to "chimney effects" (buoyancy).
    • Horizontal: Gas flows across the sample. Often better for high-gas evolution samples, but harder to load.
  • Drift: Does the balance drift over a 24-hour run? Stability is key for long isothermal tests.

C. Atmosphere Control

Can the instrument switch gases automatically?

  • Inert to Oxidative: A common TGA test involves heating in Nitrogen (to pyrolyze polymer) and then switching to Air (to burn carbon black). The gas switching manifold must be fast and automated.
  • Hermetic Sealing: For DSC, if you are measuring volatile liquids, you need a press that seals pans hermetically to withstand internal pressure (vapor).

4. The Hidden Costs: Total Cost of Ownership (TCO)

Thermal analysis is notorious for hidden costs, particularly regarding pans and cooling.

Cost Driver

Key Considerations

Sample Pans

DSC pans are consumables. Standard Aluminum pans are cheap ($1/each). High-pressure or Platinum pans are expensive ($50–$200 each). A busy lab uses thousands per year.

Purge Gas

You need dry, high-purity Nitrogen or Helium. Helium provides better heat transfer (crisper peaks) but is extremely expensive.

Cooling (LN2)

If you need sub-ambient testing, Liquid Nitrogen is a massive cost. A cooling accessory can consume 100L of LN2 per week.

Calibration

You need Indium and Zinc standards for DSC calibration. These are certified reference materials and must be replaced/recertified periodically.

5. Key Questions to Ask Vendors

Vendor specs often quote "theoretical" sensitivity that is unachievable in a real lab. Ask these practical questions to uncover the instrument's true limitations.

  1. "What is the baseline curvature (in mW) over my specific temperature range?" (Don't accept a spec for a narrow range. Ask for the full range you intend to use, e.g., -50°C to 300°C).

  2. "Does the TGA allow for automated gas switching?" (Manually swapping gas lines during a run is prone to error and introduces oxygen spikes. Integrated mass flow controllers are mandatory for reproducibility.)

  3. "Can the cooling accessory operate continuously?" (Some LN2 accessories create ice blockages if run for 24 hours. Ask about "frost-free" or heated transfer line technologies.)

  4. "Is the software capable of 'One-Click' analysis?" (For QC labs, you want a macro that automatically identifies the melting peak and calculates the area, removing operator bias.)

6. FAQ: Quick Reference for Decision Makers

Q: Do I need Modulated DSC (MDSC)?

Interested in lab tools and techniques?

Subscribe to our free Lab Tools & Techniques Newsletter.

Is the form not loading? If you use an ad blocker or browser privacy features, try turning them off and refresh the page.

By subscribing, you agree to receive email related to Lab Manager content and products. You may unsubscribe at any time.

A: If you are analyzing complex blends, curing epoxies, or trying to find a weak Tg in a crystalline polymer, yes. MDSC separates overlapping events that look like a mess in standard DSC. For simple melting points, standard DSC is sufficient.

Q: Can I run TGA and DSC on the same sample?

A: Yes, if you buy an STA (Simultaneous Thermal Analyzer). However, STA units are often compromises. They are rarely as sensitive as a dedicated DSC or as stable as a dedicated TGA. Buy an STA only if you have limited bench space or la imited sample quantity.

Q: Why does my Tg shift every time I measure it?

A: The Glass Transition (Tg) is history-dependent. If the sample was cooled quickly during manufacturing, it has a different Tg than if it were cooled slowly. To get a consistent number, you must run a "Heat-Cool-Heat" cycle: the first heat erases the thermal history, and the second heat measures the material's inherent property.

7. Emerging Trends to Watch

The field of thermal analysis is moving towards automation and hyphenation—connecting thermal instruments to chemical detectors to understand what is burning, not just that it is burning.

  • Hyphenation (TGA-IR / TGA-MS): Connecting the exhaust of a TGA to an FTIR or Mass Spec. This allows you to identify the gases evolving during decomposition. E.g., "The sample lost 10% weight, and the MS confirms that the gas was Benzene."
  • Fast-Scan DSC: New chip-based sensors allow for heating rates of 1,000°C/sec (vs. the standard 10°C/min). This allows researchers to mimic rapid industrial processing conditions (like injection molding) and suppress recrystallization.
  • Auto-Evaluation Software: AI-driven software that automatically places integration limits on peaks. This removes the "human art" of picking the start and end points of a melting peak, harmonizing results across different technicians.

Conclusion: The choice of a thermal analyzer is a choice of physical definition. DSC defines the state (solid/liquid/glass); TGA defines the composition (polymer/filler/residue); DMA defines the performance (stiff/soft/viscous). By matching the instrument physics to your material's failure mode, Lab Managers can ensure their products perform predictably in the real world.

Add Lab Manager as a preferred source on Google

Add Lab Manager as a preferred Google source to see more of our trusted coverage.

About the Author

  • Trevor Henderson headshot

    Trevor Henderson BSc (HK), MSc, PhD (c), has more than two decades of experience in the fields of scientific and technical writing, editing, and creative content creation. With academic training in the areas of human biology, physical anthropology, and community health, he has a broad skill set of both laboratory and analytical skills. Since 2013, he has been working with LabX Media Group developing content solutions that engage and inform scientists and laboratorians. He can be reached at thenderson@labmanager.com.

    View Full Profile